Bus systems
8-3
HYGROPHIL® F 5673 Operating manual, Software version 1.8.x, 387896MDHEN(14.06.2013)
Function code „03“
answers with registers of measurement variables and
parameters as showed in
Table 2
. The field 60-79 represent states and error
codes. Special measurement variables need a more detailed resolution.
Therefore some variables are spanned over two 16-bit registers (such val-
ues are marked with HI and LO in the 16-/32- bit column).
Request Host
Function code
1 byte
0x03
Start address
2 bytes
0 to 79
Number of registers
2 bytes
1 to 80
Response HYGROPHIL
®
F
Function code
1 byte
0x03
Number of bytes
1 byte
Number of transmitted registers multi-
plied with 2
Values of registers
Number of transmitted registers x 2 bytes
The values in the holding registers are spread to the available range.
A
16 bit value
has a range of values
0 ... 65535
.
Register 4, dewpoint channel 1 is between -100 °C and +100 °C.
-100 °C correlates
0
(min)
+100 °C correlates
65535
(max)
The conversion is as follows:
When HYGROPHIL for the dewpoint at channel 1 reports the register value
12345, results:
Conversion of
32 bit values
is analogous.
The range of values is here
0 ... 4294967295
.
Register value
(max – min)
65535
+ min = value
12345
100 –(-100)
65535
+ (-100) = 62,32 °C