20
Memory Setup
SP-870 provides thirty (30) user-programmable
memory test modes. To enter MEMORY
SETUP, press the SHIFT and F buttons simul-
taneously. The display should show:
Each memory comprises four items:
1.
> sign that indicates current memory mode
2.
memory mode number (01-30)
3.
wavelength
4.
factor value
Memories normally are undefined when the
instrument is delivered. An undefined memory
is followed with “None”.
1. Press the ENTER button to set up the current
memory mode that the > sign indicates.
2. Key in the desired data respectively for wave-
length and factor value, each key-in should be
followed by pressing the ENTER button.
>01 440
NM
F
XXXXXX
02 N
ONE
Note
To exit Memory Setup mode
press ESC.
value and remove the cuvette. The multiplied
Concentration value range should be 0-19999.
If this is exceeded, re-define the factor value.
11. Repeat steps 7 & 10 for further samples to be
tested at the same wavelength and factor. For
a different wavelength or factor, steps 1
through 10 should be repeated.
12. After the test is complete, turn off the power;
or, set the instrument at Standby mode if
further testing will be performed soon. (Refer
to
Standby mode section.)
O
PERATION