14
Cartridge and Membrane
Installation
Your DIamond RO has been shipped with a prefilter, car-
bon filter, MPS cartridge and one membrane (for 12 lph
unit) or two membranes (for 24 lph unit), that you will
install. To prepare your DIamond RO for operation, discon-
nect the unit from the power supply if connected.
Prefilter Installation
Particulates can damage your membranes, resulting in pre-
mature membrane failure. Therefore, your DIamond RO
uses a prefilter to remove particulates from your feed
water. Install the prefilter as follows:
1.
Open the front left door.
2.
Remove the prefilter (Part Number D502113) from
its packaging and wet the o-rings with water.
3.
Insert the prefilter straight into the left position
upper and lower quick disconnects and push until
firmly connected. For easier installation, connect
bottom first.
MPS Cartridge Installation
The MPS cartridge contains a slow disolving anti-scalant
material. The anti-scalant material combines with water to
prevent scale buildup on the membrane surface.
Install the MPS cartridge as follows:
1.
Remove the MPS cartridge (Part Number
D502114) from its packaging and wet the o-ring
with water.
2.
Insert the prefilter straight into the middle posi-
tion upper and lower quick disconnects and
push until firmly connected. For easier installa-
tion, connect bottom first.
I
NSTALLATION
Figure 6 Cartridge Installation
Note
Figure 6 shows the correct sequence of
prefilter, MPS and carbon filters.
Cartridges must be installed in the correct
order to operate properly.
Note
It will be important to replace the MPS
cartridge once the material is half dis-
solved.
Prefilter
MPS cartridge
Carbon filter