Installation / start-up
24
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe - Germany - Telephone: +49(0)5202-9801-0
Service department: +49(0)5202-9801-30 Fax: +49(0)5202-9801-99 - email: [email protected]
Instructions for Use
Barkey S-line
4.2.4
Switching off
►
Touch the ON / OFF button on the control panel if you want to switch the device
off.
INFORMATION
The device can be switched off at any time. This stops the device's operation.
4.2.5
Inserting the infusion line
We recommend that the primed infusion line is inserted in the warming section.
►
Start inserting the infusion line at the end of the warming section with the connec-
tion of the infusion line on the patient side (Fig. 9, sketches 1 and 2).
►
Insert the infusion line in the end of the warming section approx. 3 cm downstream
from the luer lock connection.
►
Move your thumb over the infusion line and press the line gently into the slot of the
warming section (Fig. 9 sketch 3).
►
Lead the infusion line to the head of the warming section (Fig. 9 sketch 4).
1.
2.
3.
4.
Fig. 9: Inserting the infusion line
For dosing clamps (roll clamps) the infusion line can be led out of the warming section at any point
or at the head or end of the device.
►
Place the head of the warming section from above into the holder for the head so
that the warming section runs downwards.
►
Connect the infusion line to the patient and secure it against tension from the
warming section.
WARNING
Make sure that the infusion line is not twisted or kinked. This can obstruct or block the infusion
flow.
The warming section may not be bent, damaged or shortened.
The warming section may not be covered, insulated, warmed or cooled.