background image

4

•  Follow all contraindications, warnings, cautions, precautions and instructions for all 

infusates, as specified by their manufacturer.

•  After use, this product may be a potential biohazard. Handle and dispose of in accor-

dance with medical practice and applicable local, state and federal laws and regulations.

•  As reported in literature, anaphylactic or anaphylactic-like reactions occur in a small 

percentage of the population during placement 

1

, positioning 

1

, flushing 

2

 of catheters, 

or cleaning of catheter exit site 

3

. These reactions are reported in association with 

insertion, rapid flushing, or manipulation of the catheter and/or use of chlorhexidine 

gluconate (CHG) in some patients. Be aware of the potential symptoms or signs of 
these reactions and take precautionary steps as dictated by institution protocol for their 

prevention or treatment.

• 

If CHG allergy is suspected, confirmatory testing is recommended 

4

,

5

•  Caution:

 Federal (USA) law restricts this device to sale by or on the order of a physician.

• 

Only qualified health care practitioners should insert, manipulate and remove these 

devices.

•  Use aseptic techniques whenever the catheter lumen is opened or connected to other 

devices. 

Precautions Related to Device Placement Procedure

•  DO NOT USE A SYRINGE SMALLER THAN 10 mL TO FLUSH AND CONFIRM PA-

TENCY. Patency should be assessed with a 10 mL syringe or larger with sterile saline. 

Upon confirmation of patency, administration of medication should be given in a syringe 

appropriately sized for the dose. Do not infuse against resistance. 

• 

To reduce potential for blood backflow into the catheter tip, always remove syringes 
slowly while injecting the last 0.5 mL of sterile saline.

•  Do not reinsert needle into IV catheter to minimize the risk of the needle damaging or 

shearing the IV catheter.

•  Do not cut guidewire to alter length.

•  The stylet or stiffening wire needs to be well behind the point the catheter is to be cut. 

NEVER cut the stylet or stiffening wire.

•  Do not insert stiff end of guidewire into vessel as this may result in vessel damage.
• 

Keep sufficient guidewire length exposed at hub to allow for proper handling. A 

non-controlled guidewire can lead to wire embolism.

•  Do not use excessive force when introducing guidewire or microintroducer as this can 

lead to vessel perforation and bleeding.

•  If the guidewire must be withdrawn while the needle is inserted, remove both needle 

and wire as a unit to prevent the needle from damaging or shearing the guidewire.

•  Never leave stylet or stiffening wire in place after catheter insertion; injury may occur. 

Remove stylet or stiffening wire and T-lock (as applicable) after insertion.

•  Do not clamp extension leg when stylet or stiffening wire is in catheter to minimize the 

risk of component or catheter damage.

•  Never use force to remove the stylet. Resistance can damage the catheter. If resistance 

or bunching of the catheter is observed, stop stylet withdrawal and allow the catheter 
to return to normal shape. Withdraw both the catheter and stylet together approximate-
ly 2 cm and reattempt stylet removal. Repeat this procedure until the stylet is easily 
removed. Once the stylet is out, advance the catheter into the desired position (zero 
mark).

Summary of Contents for PowerMidline

Page 1: ...1 Instructions For Use BARD...

Page 2: ...or past treatment will prevent proper device stabilization and or access Warnings General Warnings When using alcohol or alcohol containing antiseptics with polyurethane midlines care should be taken...

Page 3: ...dure for a particular patient A suitable trained clinician is responsible for evaluat ing the health status of a patient as it pertains to a power injection procedure Failure to warm contrast media to...

Page 4: ...the dose Do not infuse against resistance To reduce potential for blood backflow into the catheter tip always remove syringes slowly while injecting the last 0 5 mL of sterile saline Do not reinsert n...

Page 5: ...round any part of the catheter s tubing shaft or extension legs If using sutures to secure catheter USE THE SUTURE WINGS and make sure they do not occlude puncture or cut the catheter Do not use sciss...

Page 6: ...Erosion Through the Skin Catheter Embolism Catheter Occlusion Catheter Related Sepsis Exit Site Infection Exit Site Necrosis Extravasations Fibrin Sheath Formation Hematoma Heparin Induced Thrombocyto...

Page 7: ...lease tourniquet D Assess the selected vein to ensure the vessel size is adequate to accommodate the catheter being placed 2 Patient Position Catheter Measurement A Position the arm at a 90 angle B Fo...

Page 8: ...tylet assembly as one unit Caution Catheter stylet must be wetted prior to stylet repositioning or withdrawal D Retract the stylet to well behind the point the catheter is to be cut Caution Do not all...

Page 9: ...nit to prevent the needle from damaging or shearing the guidewire F Advance the small sheath and dilator together as a unit over the guidewire using a slight rotational motion If necessary a small inc...

Page 10: ...Stabilize the catheter position by applying pressure to the vein distal to the introducer sheath B Withdraw the introducer sheath from the vein and away from the site C Split the introducer sheath an...

Page 11: ...er to Instructions For Use on the proper use and removal The StatLock Stabilization Device should be monitored daily and replaced at least every seven days Caution The catheter must be secured in plac...

Page 12: ...be tween hub and wings Chevron anchor tape on top of transparent dressing Tape Strip Securement Procedure Single Lumen 1 Place 1st anchor tape over wings or bifurcation 2 Cover site and 1st anchor tap...

Page 13: ...emperature prior to power injec tion may result in catheter failure Warning Use of lumens not marked Power Injectable for power injection may result in catheter failure G Complete power injection stud...

Page 14: ...LLER THAN 10 mL TO FLUSH AND CONFIRM PATENCY Patency should be assessed with a 10 mL or larger syringe with sterile saline Upon confirmation of patency administration of medication should be given in...

Page 15: ...nd Toxic Reactions Adverse Events During Infusion Therapy Symposium University of California 1993 2 Findlay Steven R et al Hyperosmolar Triggering of Histamine Release from Human Basophils Journal of...

Page 16: ...daccess com Clinical Information 1 800 443 3385 USA Email medical services crbard com STERILIZE Manufacturer Do Not Re Use Non Pyrogenic Reorder Number Lot Number Do Not Re sterilize Length Use By Ste...

Reviews: