12
The StatLock
®
Stabilization Device Procedure
Single Lumen
1. Secure catheter with
StatLock
®
Stabilization
Device.
2. Cover site and
StatLock
®
Stabilization
Device with transpar-
ent dressing.
3. Place anchor tape
sticky side up, under
hub. Wedge tape be-
tween hub and wings.
4. Chevron anchor tape
on top of transparent
dressing.
Dual Lumen
1. Secure catheter with
StatLock
®
Stabilization
Device.
2. Cover site and
StatLock
®
Stabilization
Device with transpar-
ent dressing.
3. Place 1st anchor tape
sticky side up, under
one extension leg.
Wedge tape between
hub and wings.
Chevron anchor tape
on top of transparent
dressing.
4. Place 2nd anchor tape
sticky side up under
hub. Wedge tape be-
tween hub and wings.
Chevron anchor tape
on top of transparent
dressing.
Tape Strip Securement Procedure
Single Lumen
1. Place 1st anchor
tape over wings or
bifurcation.
2. Cover site and 1st
anchor tape with trans-
parent dressing up to
hub, but not over hub.
3. Place 2nd anchor
tape sticky side up
under hub and close to
transparent dressing.
Wedge tape between
hub and wings.
4. Chevron 2nd anchor
tape on top of trans-
parent dressing and
place 3rd anchor tape
over hub.
Dual Lumen
1. Place 1st anchor
tape over wings or
bifurcation.
2. Cover site and 1st
anchor tape with trans-
parent dressing up to
hub, but not over hub.
3. Place 2nd anchor
tape sticky side up
under hub and close to
transparent dressing.
Wedge tape between
hub and wings. Anchor
only one hub of dual
lumen catheter.
4. Chevron 2nd anchor
tape on top of trans-
parent dressing and
place 3rd anchor tape
over hub.