53
R5915130 /05
Nio 2MP / 3MP LED Display
Immunity test
IEC 60601-1-2 test levels Compliance level
Electromagnetic
environment – guidance
IEC 61000-4-8
typical location in a typical
commercial or hospital
environment
Conducted RF
IEC 61000-4-6
3 Vrms (6 Vrms in ISM
bands)
150 kHz to 80 MHz
3 Vrms (6 Vrms in ISM
bands)
-
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
Immunity to RF wireless communications equipment
Test
frequency
(MHz)
Band (MHz)
Service
Modulation
Maximum
power (W)
Distance
(m)
Immunity
test level (V/
m)
385
380 – 390
TETRA 400
Pulse
modulation
18 Hz
1.8
0.3
27
450
430 – 470
GMRS 460,
FRS 460
FM ± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704 – 787
LTE Band
13, 17
Pulse
modulation
217 Hz
0.2
0.3
9
745
780
810
800 – 960
GSM 800/
900, TETRA
800, iDEN
820, CDMA
850, LTE
Band 5
Pulse
modulation
18 Hz
2
0.3
28
870
930
1720
1700 – 1990
GSM 1800,
CDMA 1900,
GSM 1900,
DECT, LTE
Band 1/3/4/
25, UMTS
Pulse
modulation
217 Hz
2
0.3
28
1845
1970
2450
2400 – 2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
217 Hz
2
0.3
28
5240
5100 – 5800
W LAN
802.11 a/n
Pulse
modulation
217 Hz
0.2
0.3
9
5500
5785
7.6 Explanation of symbols
Symbols on the device
On the device or power supply, you may find the following symbols (nonrestrictive list):
Summary of Contents for Nio MDNC-3521 SNES
Page 20: ...R5915130 05 Nio 2MP 3MP LED Display 20 Installation and setup...
Page 37: ...37 R5915130 05 Nio 2MP 3MP LED Display Cleaning the display 5...
Page 39: ...39 R5915130 05 Nio 2MP 3MP LED Display Repackaging instructions 6...
Page 44: ...R5915130 05 Nio 2MP 3MP LED Display 44 Repackaging instructions...
Page 64: ...R5915130 05 Nio 2MP 3MP LED Display 64 Important information...
Page 65: ......
Page 66: ...www barco com R5915130 05 2023 08 22 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...