6. Important information
Power connection – Equipment with external 24 VDC power supply
•
Power requirements: The equipment must be powered using the delivered medical approved 24 VDC
(
) SELV power supply.
•
The medical approved DC (
) power supply must be powered by the AC mains voltage.
•
The power supply is speci
fi
ed as a part of the ME equipment or combination is speci
fi
ed as a ME
system.
•
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with
protective earth.
•
The equipment should be installed near an easily accessible outlet.
•
The equipment is intended for continuous operation.
•
The compliance of this monitor with Medical Safety and EMC requirements has been evaluated using
the external (optional) Skynet medical power supply model BAR-A159. If a different power supply will
be used, further investigation for Safety and EMC requirements, have to be performed at system level.
•
The compliance of this monitor with Medical Safety and EMC requirements has been evaluated using
the medical power supply and cables provided.
Power cords:
•
Utilize a UL-listed detachable power cord, 3-wire, type SJ or equivalent, 18 AWG min., rated 250 V
min., provided with a hospital-grade type plug 5-15P con
fi
guration for 120V application, or 6-15P for
240V application.
•
Do not overload wall outlets and extension cords as this may result in
fi
re or electric shock.
•
Mains lead protection (U.S.: Power cord): Power cords should be routed so that they are not likely to
be walked upon or pinched by items placed upon or against them, paying particular attention to cords
at plugs and receptacles.
•
Use a power cord that matches the voltage of the power outlet, which has been approved and complies
with the safety standard of your particular country.
Transient over-voltage
If the device is not used for a long time, disconnect it from the AC inlet to avoid damage by transient
over-voltage.
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
Connections
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1
3rd ed. or Table BBB.201 of IEC 60601-1-1 for the medical electrical systems.
Water and moisture
The equipment is IP21 compliant (IP45 front side only).
The power supply is not approved for IP21. The power supply must be mounted in a
fl
at
position for best resistance to
fl
uids.
Ventilation
Do not cover or block any ventilation openings in the cover of the set. When installing the device in a
cupboard or another enclosed location, heed the necessary space between the set and the sides of the
cupboard.
52
K5902135 (451920612561) MDSC-2326 28/02/2017
Summary of Contents for MDSC-2326 DDIH
Page 1: ...MDSC 2326 User Guide MDSC 2326 DDIH MDSC 2326 MNAH K5902135 451920612561 00 28 02 2017...
Page 16: ...2 Product overview 14 K5902135 451920612561 MDSC 2326 28 02 2017...
Page 24: ...3 Display installation 22 K5902135 451920612561 MDSC 2326 28 02 2017...
Page 34: ...4 Daily operation 32 K5902135 451920612561 MDSC 2326 28 02 2017...
Page 52: ...5 Advanced operation 50 K5902135 451920612561 MDSC 2326 28 02 2017...