6.1 Safety information
General recommendations
Read the safety and operating instructions before operating the device.
Retain safety and operating instructions for future reference.
Adhere to all warnings on the device and in the operating instructions manual.
Follow all instructions for operation and use.
Electrical Shock or Fire Hazard
To prevent electric shock or fire hazard, do not remove cover.
No serviceable parts inside. Refer servicing to qualified personnel.
Do not expose this apparatus to rain or moisture.
Modifications to the unit
Do not modify this equipment without authorization of the manufacturer.
Preventive maintenance
Performance of preventive maintenance is not essential. Periodic maintenance inspections are essential to
keep the monitor in optimum condition and ensure safe operation. We recommend a functional and safety test
of the monitor at regular intervals (e.g. at least once a year).
Type of protection (Electrical)
Equipment with external power supply: Class I equipment
Degree of safety (flammable anesthetic mixture)
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
Non-patient care equipment
•
Equipment primarily for use in a health care facility that is intended for use where contact with a patient is
unlikely (no applied part).
•
The equipment shall not be used with life support equipment.
•
The user should not touch the equipment, nor its signal input ports (SIP)/signal output ports (SOP) and the
patient at the same time.
Mission critical applications
We strongly recommend there is a replacement display immediately available in mission critical applications.
Use of Electrical Surgical Knives
Provide as much distance as possible between the electrosurgical generator and other electronic equipment
(such as monitors). An activated electrosurgical generator may cause interference with them. The interference
can activate the OSD menu of the display and as such disrupt the functionality of the display.
Power connection
–
Equipment with external 24 VDC power supply
•
Power requirements: The equipment must be powered using the delivered medical approved 24 VDC (
) SELV power supply.
•
The medical approved DC (
) power supply must be powered by the AC mains voltage.
•
The power supply is specified as a part of the ME equipment or combination is specified as a ME system.
•
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective
earth.
•
The equipment should be installed near an easily accessible outlet.
•
The equipment is intended for continuous operation.
Summary of Contents for MDSC-2324 DDI
Page 1: ...ENABLING BRIGHT OUTCOMES MDSC 2324 LED MDSC 2324 DDI MDSC 2324 MNA User Guide MDSC 2324 ...
Page 6: ...K5903252 451920612633 02 MDSC 2324 6 ...
Page 7: ...7 K5903252 451920612633 02 MDSC 2324 Welcome 1 ...
Page 10: ...K5903252 451920612633 02 MDSC 2324 10 Welcome ...
Page 11: ...11 K5903252 451920612633 02 MDSC 2324 Product overview 2 ...
Page 20: ...K5903252 451920612633 02 MDSC 2324 20 Product overview ...
Page 21: ...21 K5903252 451920612633 02 MDSC 2324 Display installation 3 ...
Page 30: ...K5903252 451920612633 02 MDSC 2324 30 Display installation ...
Page 31: ...31 K5903252 451920612633 02 MDSC 2324 Daily operation 4 ...
Page 41: ...41 K5903252 451920612633 02 MDSC 2324 Advanced operation 5 ...
Page 59: ...59 K5903252 451920612633 02 MDSC 2324 Important information 6 ...