K5903252 (451920612633) /02
MDSC-2324
28
Protective earth
Earth the MDSC-2324 by connecting the protective earth pin to a grounded outlet by means of a yellow/green
AWG18 wire (maximum admitted cable length according to national regulation requirements).
Image 3-10
CAUTION:
The display must be earthed.
Potential equalization
When potential equalization between the display and other devices is required then connect the potential
equalization pin (POAG: DIN42801) to the potential equalization terminal of the equipment.
Image 3-11
3.7 Cable routing
To route the cables
1.
For displays mounted to a VESA arm with internal cable routing provisions, route all cables through the cable
routing channel, then reinstall the connector compartment cover.
8a
Image 3-12
or
For all other mounting options, remove the cable routing channel expansion clip from the connector
compartment cover and route all cables through it while reinstalling the cover.
Summary of Contents for MDSC-2324 DDI
Page 1: ...ENABLING BRIGHT OUTCOMES MDSC 2324 LED MDSC 2324 DDI MDSC 2324 MNA User Guide MDSC 2324 ...
Page 6: ...K5903252 451920612633 02 MDSC 2324 6 ...
Page 7: ...7 K5903252 451920612633 02 MDSC 2324 Welcome 1 ...
Page 10: ...K5903252 451920612633 02 MDSC 2324 10 Welcome ...
Page 11: ...11 K5903252 451920612633 02 MDSC 2324 Product overview 2 ...
Page 20: ...K5903252 451920612633 02 MDSC 2324 20 Product overview ...
Page 21: ...21 K5903252 451920612633 02 MDSC 2324 Display installation 3 ...
Page 30: ...K5903252 451920612633 02 MDSC 2324 30 Display installation ...
Page 31: ...31 K5903252 451920612633 02 MDSC 2324 Daily operation 4 ...
Page 41: ...41 K5903252 451920612633 02 MDSC 2324 Advanced operation 5 ...
Page 59: ...59 K5903252 451920612633 02 MDSC 2324 Important information 6 ...