8. Important information
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment –
guidance
d = 1.2
√
P
80 MHz to 800
MHz
d = 2.3
√
P
800 MHz to 2.5
Ghz
Where P is the maximum
output power rating
of the transmitter in
watts (W) according
to the transmitter
manufacturer and d
is the recommended
separation distance in
meters (m).
Field strengths from
fi
xed RF transmitters,
as determined by an
electromagnetic site
survey,
7
should be less
than the compliance
level in each frequency
range.
8
Interference may occur in
the vicinity of equipment
marked with symbol:
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and re
fl
ection from structures, objects and people.
Recommended separation distance
The Nio 3MP LED Display is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer of the user of the Nio 3MP LED Display can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF com-
munications equipment (transmitters) and the Nio 3MP LED Display as recommended below, according
to the maximum output power of the communications equipment.
7. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Nio 3MP LED Display is used exceeds the applicable RF compliance level above, the Nio 3MP LED Display
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the Nio 3MP LED Display.
8. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
K5902114 NIO 3MP LED DISPLAY 03/10/2016
43
Summary of Contents for MDNC-3421
Page 1: ...Nio 3MP LED Display User Guide MDNC 3421 K5902114 01 03 10 2016...
Page 10: ...2 Parts controls and connectors 8 K5902114 NIO 3MP LED DISPLAY 03 10 2016...
Page 18: ...3 Display installation 16 K5902114 NIO 3MP LED DISPLAY 03 10 2016...
Page 22: ...4 Daily operation Cancel Standby IEC 60417 5009 20 K5902114 NIO 3MP LED DISPLAY 03 10 2016...
Page 34: ...5 Advanced operation 32 K5902114 NIO 3MP LED DISPLAY 03 10 2016...
Page 38: ...7 Cleaning your display 36 K5902114 NIO 3MP LED DISPLAY 03 10 2016...