5. Advanced operation
2. Navigate to the
Con
fi
guration > Calibration > Ambient Light
menu.
3. Enter the
Reading Room
submenu.
4. Select one of the available reading rooms and con
fi
rm.
5.14 Embedded QA
Overview
•
•
•
•
•
•
5.14.1 About Embedded QA
About
Embedded QA allows you to run a display calibration or compliance test directly from the display using the
OSD menus described in the next sections. Embedded QA will use the front sensor / I-Guard to measure
the necessary luminance levels for either a calibration or compliance test. Various settings for both actions
can be selected from the display’s OSD menu. The last results of both actions can be consulted from the
OSD.
Embedded QA or MediCal QAWeb?
Embedded QA is not a replacement for the Barco MediCal QAWeb solution.
Although Embedded QA is a reliable option to perform a simple calibration or compliance test, Barco still
highly recommends MediCal QAWeb as the solution of choice for calibration and QA. Medical QAWeb
brings many bene
fi
ts such as centralized asset management, the ability to schedule tasks, remote man-
agement, automated reporting, alerting and speci
fi
c support of regional QA standards such as DIN 6868-
57, JESRA and AAPM TG18. That’s why MediCal QAWeb Agent acts as the master for all supported dis-
plays from the moment it is installed and running. MediCal QAWeb Agent will take over from Embedded
QA and overwrite any settings which were applied by Embedded QA.
5.14.2 DICOM status report
About DICOM status report
Following information is available:
DICOM Compliance Status (status since last compliance check):
•
Compliance status:
Shows if the current DICOM curve is compliant or not.
•
Maximum error:
Shows the maximum error of the current DICOM curve. This is the deviation com-
pared to a perfect DICOM.
•
Error threshold:
Shows the error threshold. This is the maximum error allowed before a DICOM
calibration is required.
•
Time elapsed since latest compliance check:
Shows the backlight runtime since last compliance
check.
•
Display Function:
Shows the current display function.
•
Ambient light compensation:
Shows the ambient light compensation status.
•
Reading Room:
Shows the selected reading room.
•
Luminance:
Shows the measured luminance.
K5902114 NIO 3MP LED DISPLAY 03/10/2016
27
Summary of Contents for MDNC-3421
Page 1: ...Nio 3MP LED Display User Guide MDNC 3421 K5902114 01 03 10 2016...
Page 10: ...2 Parts controls and connectors 8 K5902114 NIO 3MP LED DISPLAY 03 10 2016...
Page 18: ...3 Display installation 16 K5902114 NIO 3MP LED DISPLAY 03 10 2016...
Page 22: ...4 Daily operation Cancel Standby IEC 60417 5009 20 K5902114 NIO 3MP LED DISPLAY 03 10 2016...
Page 34: ...5 Advanced operation 32 K5902114 NIO 3MP LED DISPLAY 03 10 2016...
Page 38: ...7 Cleaning your display 36 K5902114 NIO 3MP LED DISPLAY 03 10 2016...