K5902108 /06
Coronis Fusion 4MP / 6MP
49
Emissions test
Compliance
Electromagnetic environment
–
Guidance
RF emissions
CISPR 11
Group 1
The Coronis Fusion 4MP / 6MP
uses RF energy only for its internal
function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Coronis Fusion 4MP / 6MP is
suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class D
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Complies
This Coronis Fusion 4MP / 6MP complies with appropriate medical EMC standards on emissions to, and
interference from surrounding equipment. Operation is subject to the following two conditions: (1) this device
may not cause harmful interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation.
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:
•
Reorient or relocate the receiving antenna or equipment.
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The Coronis Fusion 4MP / 6MP is intended for use in the electromagnetic environment specified below. The
customer or the user of the Coronis Fusion 4MP / 6MP should assure that it is used in such an environment.
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment
–
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the Coronis Fusion 4MP /
6MP requires continued
operation during power
mains interruptions, it is
recommended that the
Summary of Contents for MDCC-4330
Page 1: ...ENABLING BRIGHT OUTCOMES MDCC 4330 MDCC 6430 User Guide Coronis Fusion 4MP 6MP...
Page 5: ...5 K5902108 06 Coronis Fusion 4MP 6MP Welcome 1...
Page 8: ...K5902108 06 Coronis Fusion 4MP 6MP 8 Welcome...
Page 18: ...K5902108 06 Coronis Fusion 4MP 6MP 18 Display installation...
Page 19: ...19 K5902108 06 Coronis Fusion 4MP 6MP Daily operation 3...
Page 23: ...23 K5902108 06 Coronis Fusion 4MP 6MP Advanced operation 4...
Page 36: ...K5902108 06 Coronis Fusion 4MP 6MP 36 Advanced operation...
Page 37: ...37 K5902108 06 Coronis Fusion 4MP 6MP Cleaning your display 5...
Page 39: ...39 K5902108 06 Coronis Fusion 4MP 6MP Repackaging instructions 6...
Page 42: ...K5902108 06 Coronis Fusion 4MP 6MP 42 Repackaging instructions...
Page 43: ...43 K5902108 06 Coronis Fusion 4MP 6MP Important information 7...