6. Important information
Immunity test
IEC 60601-1-2 4
th
edition (2014)
Test levels
Compliance level
Electromagnetic
environment –
guidance
as determined by an
electromagnetic site
survey,
2
should be less
than the compliance
level in each frequency
range.
3
Interference may occur in
the vicinity of equipment
marked with symbol:
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and re
fl
ection from structures, objects and people.
Recommended separation distance
The MDSC-2326 is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer of the user of the MDSC-2326 can help prevent electromagnetic interfer-
ence by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the MDSC-2326 as recommended below, according to the maximum output power of
the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum output
power of transmitter
4
W
150kHz to 80MHz
d=1.2
√
P
80MHz to 800MHz
d=1.2
√
P
800MHz to 2.5GHz
d=2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
2. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the MDSC-2326 is used exceeds the applicable RF compliance level above, the MDSC-2326 should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the MDSC-2326.
3. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
4. For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter. Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
60
K5902135 (451920612562) MDSC-2326 01/12/2017
Summary of Contents for High Bright MDSC-2326 DDIH
Page 1: ...MDSC 2326 User Guide MDSC 2326 DDIH MDSC 2326 MNAH K5902135 451920612562 01 01 12 2017...
Page 16: ...2 Product overview 14 K5902135 451920612562 MDSC 2326 01 12 2017...
Page 24: ...3 Display installation 22 K5902135 451920612562 MDSC 2326 01 12 2017...
Page 34: ...4 Daily operation 32 K5902135 451920612562 MDSC 2326 01 12 2017...
Page 52: ...5 Advanced operation 50 K5902135 451920612562 MDSC 2326 01 12 2017...