Cleaning
E-2320 C
61
Cleaning
Precautions
•
Take care not to damage or scratch the glass or LCD panel.
•
Do not apply pressure on the glass or LCD panel.
•
Do not apply or spray liquid directly to the glass, panel or
cabinet as excess liquid may cause damage to internal
electronics. Instead, apply the liquid to the cleaning cloth.
•
DO NOT USE:
- Lye or cleaning solutions containing lye*
- Acid
- Detergents with fluoride
- Detergents with ammonia
- Detergents with abrasives
- Steel wool
- Sponge with abrasives
- Cloth with thread made of steel
- Other coarse tools
*(Lye is a strong caustic alkaline solution of potassium salts.)
Proceed as follows:
•
Clean the cabinet using a soft cotton cloth, lightly moistened
with a recognized cleaning product for medical equipment.
•
Repeat with water only.
•
Wipe dry with a dry cloth.
•
The cabinet has been tested for resistance to the following
products:
Cidex, Betadine, Alcohol (Isopropyl and Ethyl), Ammonia-
based cleaners (Windex) and Aquasonic Gel.