The American Association of Physicists in Medicine (AAPM) composed a
list of pre-defined reading rooms. Each of these reading rooms are
defined by following parameters:
•
the maximum light allowed in this type of room
•
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset
ambient light correction value to take into account to recalculate the
DICOM display function when Ambient Light Compensation (ALC) is
enabled. Please refer to
“Ambient Light Compensation (ALC)”
page 33 to enable ALC.
The available reading rooms for your C
ORONIS
F
USION
10MP display are:
Reading rooms can only be selected when the DICOM display
function is selected. Therefore, please refer to
on page 32 to correctly set the display function.
CR/DR/MAMMO
Corresponds to light conditions in
diagnostic reading rooms for computed
radiology, digital radiology or
mammography. This setting has the lowest
maximum ambient light.
CT/MR/NM
Corresponds to light conditions in
diagnostic reading rooms for computed
tomography, magnetic resonance or
nuclear medicine scans.
Staff Office
Corresponds to light conditions in office
rooms.
Clinical Viewing
Room
Corresponds to light conditions in
diagnostic reading rooms for clinical
viewing.
Summary of Contents for Coronis Fusion 10MP
Page 1: ...User Guide Coronis Fusion 10MP...
Page 2: ...This page intentionally left blank 2 Coronis Fusion 10MP...
Page 4: ...This page intentionally left blank 4 Coronis Fusion 10MP...
Page 8: ...This page intentionally left blank 8 Coronis Fusion 10MP...
Page 44: ...This page intentionally left blank 44 Coronis Fusion 10MP...
Page 76: ...76 Coronis Fusion 10MP...
Page 77: ...Coronis Fusion 10MP 77...
Page 78: ...78 Coronis Fusion 10MP...
Page 79: ...Coronis Fusion 10MP 79...