14
9.2 Notes
9.2.1 [Note 1] Adjust the water control switch according to the direction
shown control the water volume
9.2.2 [Note 2] If the scaling tip has been screwed on tightly and there is
fine spray too, the following phenomena show that the scaling tip is damaged:
①
The vibrating intensity and the water atomization degree become weak
obviously.
②
During treatment, it produces the sound like
"
buzz
"
from the scaling tip.
9.3 If not use for a long time, please make the machine get through the
electricity and water once per month for five to ten minutes.
10 Storage and transportation
10.1 Environmental conditions of storage and transportation:
a) Relative humidity: 0 to 80%
b) Atmospheric pressure: 50kPa to 106kPa
c) Environmental temperature: -10
℃
to +50
℃
10.2 Prevent excessive shock and vibration in transportation, be sure to
handle with care and avoid inversion.
10.3 Don't mix with dangerous goods during transportation.
10.4 Avoid the sun, rain or snow during transportation.
10.5 The equipment should be handled carefully and lightly. Be sure that
it's far from the vibration, and install or store in a cool, dry and ventilated place.
10.6 Don't store the machine with the articles that are combustible,
poisonous, caustic and explosive.
Summary of Contents for P9-2
Page 1: ...P9 2 P9 3 成型尺寸 140 210MM 80g双胶纸 双色印刷 粘胶成本 ...
Page 2: ...This manual is suitable for P9 2 P9 3 series ultrasonic scaler P9 2 P9 3 ...
Page 9: ...MAX Power switch 2 6 Picture10 Picture11 Power knob Picture12 Water control switch Picture13 ...
Page 14: ...11 The circles should be replaced once they are broken or over abrasive ...
Page 19: ......
Page 20: ...PA P9 M E 1 Rev A6 ...