9.1 Troubleshooting and notes

9   Maintenance

Fault

Solutions

Possible cause

The scaling tip 
doesn't vibrate 
and there is no 
water flowing 
out when 
stepping on the 
switch

The foot switch
is in loose contact

Insert the foot switch
to its socket tightly

The fuse in the main unit  
is broken 

Contact our dealers or us

The scaling tip 
doesn't vibrate but 
there is water 
flowing out when 
stepping on the 
switch

The tip is in loose  contact

Screw the tip on the handpiece 
tightly (Picture 6 ) 

The connect plug between the
 handpiece and the circuit board
 is in loose contact

Contact our dealers or us

Contact our dealers or us

Problem of  handpiece

The scaling tip 
vibrates but there is 
no spray when 
stepping on the 
switch 

The water control switch is 
not on 

Turn on the water control switch

Note 1

 

The handpiece 
generates heat 

The water control switch
 is in a low grade

Turn the water control switch to 
a higher grade

Note 1

The vibration of 
the tip becomes
 weak 

The joint of the 
handpiece and cable 
has water leakage

The endo file doesn

'

t

vibrate

There is noise coming 
 from the endo chuck

The tip hasn't been screwed 
onto the handpiece tightly 

Screw the tip on the handpiece 
tightly(as showed in Picture 6) 

The tip is loose because
 of vibration 

"

O

is broken

The screw is loosen

The screw is loosen

The endo chuck is damage

Screw on the tip tightly
(as showed in Picture 6) 

The tip is damaged

Note 2

Change a new one

Change a new one

Change a new one

Tighten it(as showed in Picture 4) 

Tighten it(as showed in Picture 4) 

If the problem still can't be solved, please contact with local dealer or 
manufacturer.

13

There is air in the water pipe

Step on the foot switch, disassemble 
the water tank and assemble it.
 (Picture 3)

Summary of Contents for P9-2

Page 1: ...P9 2 P9 3 成型尺寸 140 210MM 80g双胶纸 双色印刷 粘胶成本 ...

Page 2: ...This manual is suitable for P9 2 P9 3 series ultrasonic scaler P9 2 P9 3 ...

Page 3: ... structure scope of application and contraindication 4 Components 5 Technical specifications 6 Usage 7 Sterilization 8 Precautions 9 Maintenance 10 Storage and transportation 11 After service 12 Environmental protections 13 Declaration Contents 1 2 7 7 8 9 10 12 13 15 14 15 15 ...

Page 4: ...Ⅱ equipment Use indoor only Type applied part Alternating current Foot switch Water control switch Follow instructions for use Ordinary equipment Drop proof Date of manufacture Appliance compliance with WEEE directive Autoclavable SN Serial number Manufacture ...

Page 5: ...e Foot switch Bottle Cover of the bottle Power plug Water filter 2 2 Accessories 2 Acquaintance of the product and its parts 2 1 The top and bottom map of main unit Scaling tips Endo chuck Picture2 1 Applied parts Endo file Picture2 2 Applied parts D2 handpiece LED or D3 handpiece light guide ...

Page 6: ...connect handpiece Sketch map for how to install the bottle D2 D3 Picture3 Picture4 4 Assemble the water bottle to the machine 1 Turn the water control switch to the maximum 2 Disassemble the handpiece 3 Make sure the power supply is connected correctly and the power switch is turned on and then step on the foot switch ...

Page 7: ...map for how to replace water filter Picture6 Sketch map for how to install tip Tighten Note The tip is tightened when the wrench slips for 2 circles Sketch map for how to install endo file Loosen Tighten Tighten Picture7 ...

Page 8: ...5 Picture9 Align O ring light guide LED lower shell D3 handpiece 2 4 Sketch map of assembling D3 Picture8 2 5 Sketch map of assembling bottle upper shell ...

Page 9: ...MAX Power switch 2 6 Picture10 Picture11 Power knob Picture12 Water control switch Picture13 ...

Page 10: ... forbidden to use this equipment 3 3 2 The patients or doctors with heart pacemaker are forbidden to use this equipment 3 3 3 The heart disease patients pregnant women and children should be cautious to use the equipment 4 Components Number Description Main unit Type Foot switch Handpiece Scaling tip Scaling tip Scaling tip Scaling tip Scaling tip Endo file Endo chuck 01 02 03 04 P9 2 P9 3 F1 S1 D...

Page 11: ... application in the presence of a Flammable Anesthetic Mixture with air or with Oxygen or Nitrous Oxide Equipment not suitable for being used in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide Continuous operation 3W to 20W 2 5kg 2N B F Ordinary equipment IPX0 Foot switch IPX1 Output primary tip vibration excursion Output tip vibration frequency Output half ...

Page 12: ... and can not be higher than the upper limit to insure safety Install the cover Picture 9 6 5 Connect the handpiece with the cable Picture 4 Hold the handpiece in the same gesture as a pen in hand 6 6 Select a suitable scaling tip as you need and screw it on the handpiece Picture 6 6 7 Connect the power and press the power switch The power is connected when it shows blue color Picture 10 6 8 Adjust...

Page 13: ...so as to avoid damage of both teeth and tip The working frequency is extremely high during operation A light touch and a certain to and fro motion will be able to eliminate the tartar without heating Overexertion and long time lingering are forbidden 6 11 2 Endo function Picture 10 Adjust the power knob until the green indicator lights Screw the endo chuck on the handpiece with endo wrench Loosen ...

Page 14: ...11 The circles should be replaced once they are broken or over abrasive ...

Page 15: ...y 8 11 Choosing the liquid according to clinic treatment 8 12 Don t pull the handpiece forcibly during operation in case damage the handpiece or the cable 8 13 Don t knock or rub the handpiece 8 14Turn off the switch and cut off the power source after stopping operation 8 15Put the liquid into the bottle and close it with its cover after the bottle is properly installed on the main unit 8 16 Only ...

Page 16: ... on the water control switch Note 1 The handpiece generates heat The water control switch is in a low grade Turn the water control switch to a higher grade Note 1 The vibration of the tip becomes weak The joint of the handpiece and cable has water leakage The endo file doesn t vibrate There is noise coming from the endo chuck The tip hasn t been screwed onto the handpiece tightly Screw the tip on ...

Page 17: ...once per month for five to ten minutes 10 Storage and transportation 10 1 Environmental conditions of storage and transportation a Relative humidity 0 to 80 b Atmospheric pressure 50kPa to 106kPa c Environmental temperature 10 to 50 10 2 Prevent excessive shock and vibration in transportation be sure to handle with care and avoid inversion 10 3 Don t mix with dangerous goods during transportation ...

Page 18: ...s been tested and homologated in accordance with EN 60601 1 2 for EMC This does not guarantee in any way that this device will not be effected by electromagnetic interference Avoid using the device in high electromagnetic environment ...

Page 19: ......

Page 20: ...PA P9 M E 1 Rev A6 ...

Reviews: