8.Operation
8.1 Open the packing and check if all the parts and
accessories are complete according to the packing list.
8
.
2
Ensure protection measures before use.For example,
wear anti-glare glasses.
8
.
3
Pull out of the lamp cap and pack it in sterilization
pouch or sterilization box to sterilize it in autoclave.
8
.
4
Install the lamp cap on the main unit.
8
.
5
Press
'
M
'
button and choose high power mode or low
power mode.
8
.
6
Press 'T' button and choose time setting.
8.7 Put the light source alignment resin which needs to be
solid, press 'P' button, and then there is a 'Di' voice. The
main unit shines blue light and starts countdown.
8
.
8
It can be stopped working at any time by short
pressing 'P' button while operating.
The next working cyble should be started after a working
8.9
cycle ends for 15 seconds in order to avoid discomfort or
scald to the patient due to the overheated lamp cap
.
8
.
10
The lamp cap gets a little hot when working normally.
But if the main unit or the lamp cap gets hot obviously,
please turn off the device until the main unit becomes cool.
If there is resin on the lamp cap after using,please
8
.
11
clean it by cotton cloth to make sure the power intensity
is good.
8
.
12
After using curing light every time, pull out of the
lamp cap, clean it and then pack it in sterilization pouch
or sterilization box to sterilize it in case using it next time.
5