background image

Hazardous Substance Statement 

Bante Instruments is committed to the reduction and eventual elimination of all hazardous substances in both the manufacturing process and 
finished products we supply. We have an active manufacturing and procurement program to minimize and eliminate the use of harmful heavy 
metals such as cadmium, lead, mercury and the like. New technologies and design parameters are also promoting these efforts and we expect 
to have little or no such materials in our product in the coming years. We welcome our customer suggestions on how to speed up these efforts. 
 

 

 
 

Warranty 

The warranty period for meter is two years from the date of shipment. Above warranty does not cover the calibration standards. Out of warranty 
products will be repaired on a charged basis. The warranty on your meter shall not apply to defects resulting from: 

•    Improper or inadequate maintenance by customer.

 

•    Unauthorized modification or misuse.

 

•    Operation outside of the environment specifications of the products.

 

For more information, please contact the supplier. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
Bante Instruments Inc. 
4715 Castlewood St., Sugar land, TX 77479, USA 
E-mail: [email protected] 
 

The information in this document is subject to change without notice. Copyright ©  Bante Instruments, 2019. All rights reserved. 

 

Summary of Contents for TB200

Page 1: ...TB200 Laboratory Turbidity Meter Instruction Manual Bante Instruments Inc...

Page 2: ...bration 7 Turbidity measurement 9 Low turbidity samples 200 NTU 9 High turbidity samples 2000 NTU 9 Total suspended solids measurement 10 Calculating the TSS factor 10 Setting the TSS factor 10 TSS me...

Page 3: ...ent work environment meets the following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C 32 F and less than 50 C 122 F No potential magnetic fields and ambient lig...

Page 4: ...oll down through the menu item Confirms the calibration settings or displayed options Starts measurement Locks the measured value Switching the Meter On and Off Insert the connector of power adapter i...

Page 5: ...0 1 0 01 TSS Factor Set the conversion factor for measurement of the total suspended solids 0 13 Auto Power Off 2 hours When the option is enabled the meter will automatically switch off if no key is...

Page 6: ...libration or settings 3 1 Press and hold the key for 3 seconds to enter the setup menu 3 2 Press the key to select the Password press the Enter key to confirm 3 3 Press the key to select the Enable Pr...

Page 7: ...brated 4 1 Press and hold the key for 3 seconds to enter the setup menu 4 2 Press the key to select the Reset press the Enter key to confirm 4 3 Press the key to select Restore factory settings Press...

Page 8: ...ause inaccurate readings Turbidity Calibration The TB200 turbidity meter is calibrated with Formazin Standards at the factory and does not require user calibration prior to use Do not shake or agitate...

Page 9: ...th the arrow on the meter Close the light shield lid 6 Press the Enter key the meter begins the calibration The Calibrating icon shows on the upper right corner of the screen 7 Wait for the reading to...

Page 10: ...ops and fingerprints Ensure that the outside of vial is dry and clean 1 4 Place the vial in the measurement chamber Align the mark on the vial with the arrow on the meter 1 5 Close the light shield li...

Page 11: ...for them decrease to room temperature 1 8 Weigh the filter disk and watch glass and record the mg value as A 1 9 Calculate the TSS value using the formula below TSS mg L 1 10 Fill the vial with the s...

Page 12: ...y to switch pages 3 Press the ESC key the meter returns to the measurement mode Deleting the data log If the memory is full the meter will automatically show a reminder screen when the MI key is press...

Page 13: ...uter 2 Click the OK button the application starts If your computer can not find a communication port please click the file CP210xVCPInstaller_x64 exe or CP210xVCPInstaller_x86 exe to update the drive...

Page 14: ...urbidity free water The turbidity free water is used for preparation of turbidity standards and is prepared by filtering deionized water through a 0 45 m or smaller pore sized membrane Preparation of...

Page 15: ...ight Source Infrared emitting diode Detector Silicon photodiode Stray Light 0 02 NTU Sample Vial 60 H 25 Dia mm Memory Stores up to 200 data sets Output USB communication interface Display 3 5 TFT dis...

Page 16: ...r suggestions on how to speed up these efforts Warranty The warranty period for meter is two years from the date of shipment Above warranty does not cover the calibration standards Out of warranty pro...

Reviews: