background image

Measurement and Calibration

 

 

Turbidity Measurement 

An accurate turbidity measurement depends on good measurement techniques. Factors such as clean sample vials, positioning of vial in the   
measurement chamber, covering the vial with the light shield lid, meter calibration, removal of bubbles, etc. For more information, please refer to 
page 7 

Measurement and Calibration Hints

 

Low turbidity samples (< 200 NTU) 

For the low turbidity samples, we recommend using the same vial to perform the measurement and calibration. 
1.1.  Rinse the vial with approximately 10ml of the sample, capping the vial with the screw cap and gently inverting it several times. Discard 
 

the used sample and repeat the rinsing procedure two more times. 

1.2.  Fill the vial with the sample. Cap the vial. 
1.3      Wipe the vial with the lint-free cloth to remove waterdrops and fingerprints. Ensure that the outside of vial is dry and clean. 
1.4  Place the vial in the measurement chamber. Align the mark on the vial with the arrow on the meter. 
1.5  Close the light shield lid. Press the 

Meas

 key, the meter begins the measurement, the

 

 

icon shows on screen. 

1.6  Wait for the reading to stabilize, the icon will automatically switch to

1.7  Press the 

Meas

 key, the meter will take a measurement again. 

 

   

 

 

  In the measurement process, press the

key. The measured value will be locked or unlocked. 

 
High turbidity samples (>2000 NTU) 

For the high turbidity samples, the solution must be diluted prior to measurement. The dilution water can be obtained by filtering deionized water 
through a < 0.45µ m filter membrane. 
2.1  Repeat steps 1.1 through 1.5 above and record the measured value. 
2.2  Calculate the true turbidity of the original sample using the following formula. 
 
 
 

T =   

 

 

 

Where:

   

T

 

 

= Ture turbidity of the original sample 

 

   

  T

d

 = Measured value 

 

   

  V

s

 = Volume of the original sample (ml) 

 

   

  V

d

 = Volume of the dilution water (ml) 

 
 
 
 
 
 
 
 

T

d

 (V

s

 + V

d

V

s

 

Summary of Contents for TB200

Page 1: ...TB200 Laboratory Turbidity Meter Instruction Manual Bante Instruments Inc...

Page 2: ...bration 7 Turbidity measurement 9 Low turbidity samples 200 NTU 9 High turbidity samples 2000 NTU 9 Total suspended solids measurement 10 Calculating the TSS factor 10 Setting the TSS factor 10 TSS me...

Page 3: ...ent work environment meets the following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C 32 F and less than 50 C 122 F No potential magnetic fields and ambient lig...

Page 4: ...oll down through the menu item Confirms the calibration settings or displayed options Starts measurement Locks the measured value Switching the Meter On and Off Insert the connector of power adapter i...

Page 5: ...0 1 0 01 TSS Factor Set the conversion factor for measurement of the total suspended solids 0 13 Auto Power Off 2 hours When the option is enabled the meter will automatically switch off if no key is...

Page 6: ...libration or settings 3 1 Press and hold the key for 3 seconds to enter the setup menu 3 2 Press the key to select the Password press the Enter key to confirm 3 3 Press the key to select the Enable Pr...

Page 7: ...brated 4 1 Press and hold the key for 3 seconds to enter the setup menu 4 2 Press the key to select the Reset press the Enter key to confirm 4 3 Press the key to select Restore factory settings Press...

Page 8: ...ause inaccurate readings Turbidity Calibration The TB200 turbidity meter is calibrated with Formazin Standards at the factory and does not require user calibration prior to use Do not shake or agitate...

Page 9: ...th the arrow on the meter Close the light shield lid 6 Press the Enter key the meter begins the calibration The Calibrating icon shows on the upper right corner of the screen 7 Wait for the reading to...

Page 10: ...ops and fingerprints Ensure that the outside of vial is dry and clean 1 4 Place the vial in the measurement chamber Align the mark on the vial with the arrow on the meter 1 5 Close the light shield li...

Page 11: ...for them decrease to room temperature 1 8 Weigh the filter disk and watch glass and record the mg value as A 1 9 Calculate the TSS value using the formula below TSS mg L 1 10 Fill the vial with the s...

Page 12: ...y to switch pages 3 Press the ESC key the meter returns to the measurement mode Deleting the data log If the memory is full the meter will automatically show a reminder screen when the MI key is press...

Page 13: ...uter 2 Click the OK button the application starts If your computer can not find a communication port please click the file CP210xVCPInstaller_x64 exe or CP210xVCPInstaller_x86 exe to update the drive...

Page 14: ...urbidity free water The turbidity free water is used for preparation of turbidity standards and is prepared by filtering deionized water through a 0 45 m or smaller pore sized membrane Preparation of...

Page 15: ...ight Source Infrared emitting diode Detector Silicon photodiode Stray Light 0 02 NTU Sample Vial 60 H 25 Dia mm Memory Stores up to 200 data sets Output USB communication interface Display 3 5 TFT dis...

Page 16: ...r suggestions on how to speed up these efforts Warranty The warranty period for meter is two years from the date of shipment Above warranty does not cover the calibration standards Out of warranty pro...

Reviews: