background image

13 

Appendix

 

 

Appendix   

 

Indexing a sample vial 

The United States Environmental Protection Agency (U.S. EPA) recommends that the vial used for turbidity calibration or sample measurement be 
indexed. Its purpose is obtain a position which provides the lowest turbidity reading. The indexing methods are as follows. 
1.1  Fill the vial using the distilled water (< 0.5 NTU). Cap the vial. 
1.2  Wipe the vial with the lint-free cloth to remove waterdrops and fingerprints. 
1.3  Place the vial in the measurement chamber. Align the mark on the vial with the arrow on the meter. 
1.4  Press the 

Meas

 key, the meter begins the measurement. 

1.5  Slowly rotate the vial approximately 45 degrees. Close the light shield lid and record the measured value. 
1.6  Repeat the step 1.5 until the lowest turbidity reading is shown. 
1.7  Mark this position on the vial. 
 

Matching the sample vials 

For the best accuracy and repeatability, using one indexed vial is best choice for turbidity measurement. If you need to use a few vials, match   
these vials are necessary. 
2.1  Repeat the steps above for each vial and record the measured values. 
2.2  Find the closest position of these vials measuring value and mark it. 
 

Preparing the formazin standards 

Preparation of turbidity-free water: 
The turbidity-free water is used for preparation of turbidity standards and is prepared by filtering deionized water through a 0.45µ m or smaller 
pore-sized membrane. 
 
Preparation of turbidity standards: 

 

Turbidity Standard 

Preparation Method 

4000 NTU 

 

Dissolve 1g hydrazine sulfate [(NH

2

)

2

H

2

SO

4

] in the turbidity-free water and dilute to 100ml in a volumetric flask. 

 

Dissolve 10g hexamethylenetetramine [(CH

2

)

 

6

 

N

4

] in the turbidity-free water and dilute to 100ml in a volumetric

.

 

 

flask. 

 

Mix 5ml of hydrazine sulfate and 5ml of hexamethylenetetramine solutions in a 100ml volumetric flask and let

.

   

 

stand 24 hours at 25

 

(± 3)° C. 

2000 NTU 

 

Mix 50ml of 4000 NTU standard in a 100ml volumetric flask and dilute to the mark. 

1500 NTU 

 

Mix 37.5ml of 4000 NTU standard in a 100ml volumetric flask and dilute to the mark. 

1000 NTU 

 

Mix 25ml of 4000 NTU standard in a 100ml volumetric flask and dilute to the mark. 

500 NTU 

 

Mix 12.5ml of 4000 NTU standard in a 100ml volumetric flask and dilute to the mark. 

200 NTU 

 

Mix 10ml of 4000 NTU standard in a 100ml volumetric flask and dilute to the mark. 

 

Mix 50ml of above standard in a 100ml volumetric flask and dilute to the mark. 

10 NTU 

 

Mix 10ml of 4000 NTU standard in a 100ml volumetric flask and dilute to the mark. 

 

Mix 2.5ml of above standard in a 100ml volumetric flask and dilute to the mark. 

 

Or,

 

mix 2ml of 500

 

NTU standard in a

 

100ml volumetric flask and dilute to the mark. 

 

 

 

 

Summary of Contents for TB200

Page 1: ...TB200 Laboratory Turbidity Meter Instruction Manual Bante Instruments Inc...

Page 2: ...bration 7 Turbidity measurement 9 Low turbidity samples 200 NTU 9 High turbidity samples 2000 NTU 9 Total suspended solids measurement 10 Calculating the TSS factor 10 Setting the TSS factor 10 TSS me...

Page 3: ...ent work environment meets the following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C 32 F and less than 50 C 122 F No potential magnetic fields and ambient lig...

Page 4: ...oll down through the menu item Confirms the calibration settings or displayed options Starts measurement Locks the measured value Switching the Meter On and Off Insert the connector of power adapter i...

Page 5: ...0 1 0 01 TSS Factor Set the conversion factor for measurement of the total suspended solids 0 13 Auto Power Off 2 hours When the option is enabled the meter will automatically switch off if no key is...

Page 6: ...libration or settings 3 1 Press and hold the key for 3 seconds to enter the setup menu 3 2 Press the key to select the Password press the Enter key to confirm 3 3 Press the key to select the Enable Pr...

Page 7: ...brated 4 1 Press and hold the key for 3 seconds to enter the setup menu 4 2 Press the key to select the Reset press the Enter key to confirm 4 3 Press the key to select Restore factory settings Press...

Page 8: ...ause inaccurate readings Turbidity Calibration The TB200 turbidity meter is calibrated with Formazin Standards at the factory and does not require user calibration prior to use Do not shake or agitate...

Page 9: ...th the arrow on the meter Close the light shield lid 6 Press the Enter key the meter begins the calibration The Calibrating icon shows on the upper right corner of the screen 7 Wait for the reading to...

Page 10: ...ops and fingerprints Ensure that the outside of vial is dry and clean 1 4 Place the vial in the measurement chamber Align the mark on the vial with the arrow on the meter 1 5 Close the light shield li...

Page 11: ...for them decrease to room temperature 1 8 Weigh the filter disk and watch glass and record the mg value as A 1 9 Calculate the TSS value using the formula below TSS mg L 1 10 Fill the vial with the s...

Page 12: ...y to switch pages 3 Press the ESC key the meter returns to the measurement mode Deleting the data log If the memory is full the meter will automatically show a reminder screen when the MI key is press...

Page 13: ...uter 2 Click the OK button the application starts If your computer can not find a communication port please click the file CP210xVCPInstaller_x64 exe or CP210xVCPInstaller_x86 exe to update the drive...

Page 14: ...urbidity free water The turbidity free water is used for preparation of turbidity standards and is prepared by filtering deionized water through a 0 45 m or smaller pore sized membrane Preparation of...

Page 15: ...ight Source Infrared emitting diode Detector Silicon photodiode Stray Light 0 02 NTU Sample Vial 60 H 25 Dia mm Memory Stores up to 200 data sets Output USB communication interface Display 3 5 TFT dis...

Page 16: ...r suggestions on how to speed up these efforts Warranty The warranty period for meter is two years from the date of shipment Above warranty does not cover the calibration standards Out of warranty pro...

Reviews: