background image

 

 

Appendix 

 

Preparation of pH Buffer Solutions 

The PHscan40 tester is packaged with the pH

 

4.01/7.00/10.01 buffer 

sachets required for calibration. 
1.  Half fill a 250 ml volumetric flask with distilled water and add the 
 

pH 7.00 buffer reagent. 

 

2.    Swirl the volumetric flask gently to dissolve the reagent and fill 
 

to the mark with distilled water. 

 

3.    Cap and upend the volumetric flask several times to mix solution. 
 

 

  Preparation of pH 4.01 and 10.01 buffer solutions are the same 

 

as above.   

  Prepared buffer solution should be stored in hermetically sealed 

 

glass container and avoid direct sunlight. 

 

Preparation of Electrode Storage Solution 

Dissolve 24.6 grams of analytical grade KCl reagent in 100 ml distilled 
water. Add pH 4.01 standard buffer and adjust solution to pH 4. 
 

Optional Accessories 

 

pH Electrode 

 
Order Code 

Description 

E201-BNC 

General purpose pH electrode, epoxy body, 12

 

mm 

(0.47 in.) diameter. For liquids. 

E202-BNC 

Flat surface pH electrode, epoxy body, 12

 

mm   

(0.47 in.) diameter. For semisolid samples. 

P11 

General purpose pH electrode, Glass body, 12

 

mm 

(0.47 in.) diameter. For weak corrosive liquids. 

 

5   

I   

PHscan40 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

250 ml 

 

 

Electrode Maintenance 

 

Cleaning the Electrode 

Since pH electrode is susceptible to contamination, thoroughly clean 
as necessary after each use. 

  General Cleaning 

 

Rinse the pH electrode with distilled water and soak in 3M KCl 

 

solution. 

 

  Salt Deposits 

 

Dissolve the deposit by  immersing  the  electrode in  warm  tap 

 

water. Rinse the electrode with distilled water and soak in 3M

 

KCl solution. 

 

  Oil or Grease 

 

Place the electrode in the detergent or ethanol solution for 15 

 

minutes. Rinse the electrode with distilled water and soak in 3M 

 

KCl solution. 

 

  Protein 

 

(1)  Add 1% pepsin to 0.1M HCl solution. 

 

(2)  Place the electrode in above solution for 15 minutes. 

 

(3)  Rinse the electrode with distilled water and soak in 3M KCl 

 

 

solution. 

 

  Clogged Liquid Junction 

 

(1)  Heat a diluted KCl solution to 60° C (140° F). 

 

(2)  Place the electrode into the heated solution for 10 minutes. 

 

(3)  Allow the electrode to cool in unheated KCl solution. 

 

Reactivating the Electrode 

If the pH-sensitive membrane has dried out, the electrode response 
will become sluggish. Immerse the electrode in a pH 4.01 buffer   
solution for about 30 minutes to rehydrate. If this fails, the electrode 
requires activation. 
1.  Soak the electrode in 0.1M HCl for 10 minutes.   
2.  Remove  and  rinse  with  distilled  water,  then  place  into  0.1M 
 

NaOH for 10 minutes. 

3.  Remove and rinse again, and soak in 3M KCl solution for at least 
 

6 hours. 

If these steps fail to restore the response, replace the electrode. 
 

Storing the Electrode 

For best results, always soak the electrode in 3M KCl solution. If above 
solution is not available, use a pH 4.01 buffer solution. 
 

 

DO NOT store the electrode in distilled or deionized water that will 
deplete the hydration layer of the pH-sensitive membrane and render 
the electrode useless. 
If you do not use the electrode for a period longer than 1 month, store 
the electrode in storage solution. 

Summary of Contents for PHscan40

Page 1: ...PHscan40 Pocket pH Tester USER MANUAL Bante Instruments Inc ...

Page 2: ...inimum power requirements the icon automatically disappears MEAS Indicates that the tester is in the measurement mode CAL Indicates that the tester is in the calibration mode SETUP Indicates that the tester is in the setup mode ATC Indicates that the temperature compensation is enabled Prior to Use 1 Take out the pH electrode from the carrying case Insert the BNC connector into the connector socke...

Page 3: ...ble Disable default Auto Power Off If enabled the tester will automatically switch off if no key is pressed within 8 minutes Enable default Disable Factory Reset If enabled all of the calibration values and current settings will be deleted or reset to the factory defaults the tester must be recalibrated Enable Disable default Setting the Default Option 1 In the measurement mode press and hold the ...

Page 4: ...bration The PHscan40 tester allows 1 to 3 points calibration we recommend that you perform at least 2 points calibration for accurate pH measurement the tester will automatically recognize and calibrate to following standard buffer values USA standards pH 4 01 7 00 10 01 NIST standards pH 4 01 6 86 9 18 Single point calibration should only be carried out with the pH 7 00 or 6 86 otherwise calibrat...

Page 5: ...pH 10 01 buffer solution stir the electrode gently Press the Enter key to begin the calibration When the reading has stabilized the display will automatically shows the electrode slope and Calibration is completed During the calibration if the display shows indicating that the measured mV value for the current calibration point deviates by more than 60 mV approximately 1 pH from the theoretical va...

Page 6: ... deposit by immersing the electrode in warm tap water Rinse the electrode with distilled water and soak in 3M KCl solution Oil or Grease Place the electrode in the detergent or ethanol solution for 15 minutes Rinse the electrode with distilled water and soak in 3M KCl solution Protein 1 Add 1 pepsin to 0 1M HCl solution 2 Place the electrode in above solution for 15 minutes 3 Rinse the electrode w...

Page 7: ...PHscan40 Range 1 00 to 15 00 pH Resolution 0 01 pH Accuracy 0 01 pH Calibration Point 1 to 3 points pH Buffer Option USA or NIST Automatic Buffer Recognition pH4 01 7 00 10 01or4 01 6 86 9 18 Temperature Compensation 0 to 100 C 32 to 210 F manual Operating Temperature 0 to 50 C 32 to 122 F Storage Temperature 0 to 60 C 32 to 140 F Relative Humidity 80 non condensing IP Rating IP54 Display LCD 21 2...

Page 8: ...ster is one year from the date of shipment Above warranty does not cover the electrode and pH buffer solutions Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mor...

Reviews: