PHscan40 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Overview 

Thank you for selecting the PHscan series pocket pH tester. This user 
manual provides a step-by-step guide to help you operate the tester, 
please carefully read the following instructions before use. 
 

Installing the Batteries 

1.  Twist the connector collar counter clockwise, pull the BNC   
 

connector away from the tester. 

 

 

2.  Insert two AAA alkaline batteries into the battery compartment, 
 

note polarity. 

 

 

3.  Push the BNC connector into the tester and twist the connector

 

 

collar clockwise until tight. 

 

 

Keypad 

 
Key   

 

Function 

 

   

Switch the tester on or off 

   

Lock or unlock measurement 

   

Exit the calibration, settings and return to the pH 

 

   

measurement 

 

   

Start calibration 

   

Press and hold the key to enter the setup menu 

   

Select an option 

 

   

Confirm the calibration, settings or displayed option 

 

Display 

 
Icon   

 

Description 

 

When the battery voltage falls below the minimum   
power requirements, the icon automatically   
disappears 

 
 
 

 
MEAS 

Indicates that the tester is in the measurement mode 

CAL 

Indicates that the tester is in the calibration mode 

SETUP 

Indicates that the tester is in the setup mode 

ATC 

Indicates that the temperature compensation is enabled 

 

 

 

Prior to Use 

1.  Take out the pH electrode from the carrying case. Insert the BNC 
 

connector into the connector socket on tester, rotate and push 

 

the connector clockwise until it locks. 

 

 

 

 
2.  Remove the protective cap from the bottom of the electrode. If 
 

tiny air bubbles are present inside the pH-sensitive glass   

 

membrane, gently shake the electrode downward to remove air 

 

bubbles. 

 

 

 

 

 

   

 

 
3.  If the glass membrane has dried out, soak the electrode in 3M 
 

KCl or pH 4.01 buffer solution for about 30 minutes. 

 

 

 

Switching the Tester On and Off 

  Press and hold the 

Meas

 key for about 5 seconds to switch on 

 

the tester. 

  Press and hold the 

Meas

 key to switch off the tester. 

 

 

 

If you do not press any key within 8 minutes, the tester will switch off 
automatically to conserve energy. 

3M KCl solution 
 

 

Summary of Contents for PHscan40

Page 1: ...PHscan40 Pocket pH Tester USER MANUAL Bante Instruments Inc ...

Page 2: ...inimum power requirements the icon automatically disappears MEAS Indicates that the tester is in the measurement mode CAL Indicates that the tester is in the calibration mode SETUP Indicates that the tester is in the setup mode ATC Indicates that the temperature compensation is enabled Prior to Use 1 Take out the pH electrode from the carrying case Insert the BNC connector into the connector socke...

Page 3: ...ble Disable default Auto Power Off If enabled the tester will automatically switch off if no key is pressed within 8 minutes Enable default Disable Factory Reset If enabled all of the calibration values and current settings will be deleted or reset to the factory defaults the tester must be recalibrated Enable Disable default Setting the Default Option 1 In the measurement mode press and hold the ...

Page 4: ...bration The PHscan40 tester allows 1 to 3 points calibration we recommend that you perform at least 2 points calibration for accurate pH measurement the tester will automatically recognize and calibrate to following standard buffer values USA standards pH 4 01 7 00 10 01 NIST standards pH 4 01 6 86 9 18 Single point calibration should only be carried out with the pH 7 00 or 6 86 otherwise calibrat...

Page 5: ...pH 10 01 buffer solution stir the electrode gently Press the Enter key to begin the calibration When the reading has stabilized the display will automatically shows the electrode slope and Calibration is completed During the calibration if the display shows indicating that the measured mV value for the current calibration point deviates by more than 60 mV approximately 1 pH from the theoretical va...

Page 6: ... deposit by immersing the electrode in warm tap water Rinse the electrode with distilled water and soak in 3M KCl solution Oil or Grease Place the electrode in the detergent or ethanol solution for 15 minutes Rinse the electrode with distilled water and soak in 3M KCl solution Protein 1 Add 1 pepsin to 0 1M HCl solution 2 Place the electrode in above solution for 15 minutes 3 Rinse the electrode w...

Page 7: ...PHscan40 Range 1 00 to 15 00 pH Resolution 0 01 pH Accuracy 0 01 pH Calibration Point 1 to 3 points pH Buffer Option USA or NIST Automatic Buffer Recognition pH4 01 7 00 10 01or4 01 6 86 9 18 Temperature Compensation 0 to 100 C 32 to 210 F manual Operating Temperature 0 to 50 C 32 to 122 F Storage Temperature 0 to 60 C 32 to 140 F Relative Humidity 80 non condensing IP Rating IP54 Display LCD 21 2...

Page 8: ...ster is one year from the date of shipment Above warranty does not cover the electrode and pH buffer solutions Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mor...