NanoRam
®
Series
Handheld Raman
19 Shea Way, Newark, DE 19713, USA
•
Tel: +1 (302) 368-7824
•
Fax: +1 (302) 368-7830
•
Web: www.bwtek.com
290020255-J (2017/07/28)
Page 21 of 38
Copyright 2017 B&W Tek
Using the Liquid Vial Holder Adaptor
The liquid vial holder can be used for liquid or gel samples. The liquid vial holder holds a
15mm (diameter) X 45mm (height) vial.
To install liquid vial holder adaptor, insert the adaptor onto the shaft at the laser aperture.
Make sure the vial holder key is aligned with the jacket rib. Gently push it all the way until
the adaptor is now installed properly into position and secured by the rubber jacket ribs.
Using the Rubber Jacket’s Ribs as Bottle Adaptor
The rubber jacket’s rib is designed in such a shape that it can be used as a bottle adaptor
for the measurement of samples through transparent containers of less than 5mm in
thickness, such as liquid samples inside bottle containers. To scan liquid inside the bottle,
first remove any sampling adaptor. With only the shaft in place, bottle or container can be
placed directly against the shaft. The measurement beam will be ideally focused into the
sample over 5mm away from the surface of the adapter.
Practice caution when measuring dark samples as they may heat
up and potentially burn. Lower laser power settings are
recommended for dark materials.
Align the vial
holder key
with the rib