Functional check following mounting
To perform this functional check, the tank out
fl
ow must be closed.
1. Fill the tank with water up to the
fi
lling level mark.
2. Turn on the ultrasonic bath from the TRISON Base's power switch, follow the
instructions for use to do so.
3. If available connect the TRISON Twist to the control unit.
4. Select the "Robotic Instruments" program and press
until the channel selection
appears. Select both couplings (A+B) and start. Skip soaking (soaking running) by
pressing the
Button and allow program to run for approx. 5 min. No errors may be
displayed. During this time, look for any leaks in the bottom-mount cabinet. Check for
movements at the connected TRISON Twist.
5. Stop the process and turn off the ultrasonic bath once again.
6. Drain the water.
Once mounting is completed, af
fi
x the enclosed safety adhesives on clearly visible spots
near the ultrasonic oscillating tank, e.g. on the front of the bottom-mount cabinet.
9.1
Foil test
After all execute a foil test according the enclosed information.
10
Technical data
TRISON ultrasonic baths are interference-free and CE - marked.
Safety: EN 61010-1
EMC: EN 61326-1
Ambient conditions according to EN 61 010-1
- Overvoltage category:
II
- Degree of contamination:
2
- Protection class:
I
- Permissible ambient temperature:
15 to 35 °C
- Permissible relative humidity up to 31°C: 80 %
- Permissible relative humidity up to 40 °C: 50 %
- Altitude:
up to 2000 m
- No condensation allowed.
- Only for indoor operation.
“Details on use as a medical device” and “Information on EMC” can be found in the
Instructions for Use.