2× Ø 10.5, 160 mm apart
(clearance hole for
screw heads)
Width 6 mm
61279e GB/2018-03
13 / 25
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
4.5
Mounting the drain set
The drain set is already pre-mounted on the tank.
Piping with an odour trap must be properly installed in order to connect the drain set to
a nearby drainage system.
5
HF generator
5.1
Suspending the HF generator (recommended)
The HF generator is to be suspended next
to the ultrasonic oscillating tank dust-free
and dry. A minimum distance of 10 cm in all
directions is required for suf
fi
cient cooling
installation in an enclosed structure is not
permitted.
1. Drill holes for the screws (e.g. M5) leaving
160 mm of distance between them and
ensuring that there is suf
fi
cient height.
2. Attach the HF generator.
The HF generator can be suspended on either the right- or
left-hand side (supplied from the factory as a right-hand
version).
5.2
Positioning the HF generator
Position the HF generator dust-free and dry on a
fi
rm base near the ultrasonic
oscillating tank. A minimum distance of 10 cm in all directions is required for suf
fi
cient
cooling.
Installation in an enclosed structure is not permitted.
6
Control unit
Depending on the version, the TRISON Base control unit must be mounted on either
the right- or left-hand side of the work surface, next to the ultrasonic oscillating tank
(tank outlet must point to the control unit!). The information and diagrams in chapter 3
must be regarded.
The following must be taken into consideration prior to assembly and drilling of the
holes.
• The unit can be swivelled on its base by up to 180° in increments of 15° (12 stops)