Ambient conditions
Overvoltage category:
II
Degree of contamination:
1
Permissible ambient temperature:
- 20 to 40 °C
Permissible relative humidity to 31 °C: 65%
Permissible relative humidity to 40 °C: 65%
Permissible air pressure:
800 to 1050 mbar
No dewing.
For indoor operation only.
Electromagnetic compatibility (EMC)
The device has been tested for electromagnetic compatibility (EMC) and meets
the requirements for Class B devices in accordance with EN 55011.
It is suitable for use in institutions and other areas that are directly connected to
a public low-voltage power supply, e.g., medical laboratory facilities.