background image

8 / 16

51766 GB/2020-06

1

 electronic GmbH & Co. KG  •  Heinrichstraße 3-4  •  12207 Berlin  •  Deutschland  •  [email protected]

1.4  

Technical data  

The SONOPULS sound proof box LS 40 is interference-free and CE-marked.
Safety: EN 61010-1; 
EMC: EN 61326-1

Mains supply: 

100-240 V~ (± 10%) 50/60 Hz, with separate power supply

Housing: 

stainless steel AISI 304 (1.4301) 

Splash guard: 

stainless steel AISI 304 (1.4301)

Drip tray: 

stainless steel AISI 304 (1.4301)

Perforated plate: 

stainless steel AISI 304 (1.4301)

Insulation: ester-acoustic 

foam

External dimensions: 

445 × 515 × 1000 mm (L×W×H)

Internal dimensions: 

350 × 310 × 880 (L×W×H)

Noise level reduction: 

approx. 30 dB (AU)

Weight: 30 

kg

Protection class: 

III

Serial number (SN): 

see type plate

Degree of protection: 

IP 2X pursuant to EN 60529

Ambient conditions pursuant to EN 61 010-1

Overvoltage category: 

II

Degree of contamination: 

2

Permissible ambient temperature: 

5 to 40 °C

Permissible relative humidity up to 31 °C:  80%
Permissible relative humidity up to 40 °C:  50%
Altitude: 

up to 2000 m above sea level

No condensation allowed.
For indoor operation only.

1.4.1  Electromagnetic compatibility (EMC)  

The device has been checked for electromagnetic compatibility (EMC) in accordance with DIN 
EN 61326-1 and meets the requirements for Class B devices in accordance with EN 55011. 
It is suitable for use in institutions and other areas that are directly connected to a public low-
voltage power supply, e.g., medical laboratory facilities.

Protected to prevent 

fi

 ngers from 

accessing dangerous parts. 

Protected against solid foreign 

bodies with a diameter of 12 mm 

or larger.

Summary of Contents for SONOPULS LS 40

Page 1: ...7 LS 40 Sound proof box 51766 GB 2020 06 Instructions for Use...

Page 2: ...nal version will take precedence BANDELIN accepts no responsibility or liability for damage caused by improper handling or usage contrary to the intended purpose The documentation has been prepared wi...

Page 3: ...y responsibility for the device s safety or functional ability in the event of improper handling or usage contrary to the intended purpose Unauthorised alterations modifications will render both the w...

Page 4: ...1 4 Technical data 8 1 4 1 Electromagnetic compatibility EMC 8 1 5 Warnings and safety instructions 9 2 Preparation 10 2 1 Scope of delivery 10 2 2 Installation assembly 10 2 3 Commissioning 11 3 Ope...

Page 5: ...6 1 electronic GmbH Co KG Heinrichstra e 3 4 12207 Berlin Deutschland info bandelin com 6 Maintenance and repair 14 6 1 Servicing 14 6 2 Troubleshooting 14 6 3 Repairs and service 14 7 Accessories 15...

Page 6: ...pe plate on the rear Product features 1 Body 2 Door with acrylic glass window 3 Handle 4 Seal 5 Perforated plate removable 6 Drip tray removable 7 LED interior lighting 8 Ventilation system 9 Splash g...

Page 7: ...sample fluids are collected in the drip tray For a necessary cleaning the drip tray can be removed The perforated plate above the drip tray enables a stable positioning of the stand or sample vessels...

Page 8: ...type plate Degree of protection IP 2X pursuant to EN 60529 Ambient conditions pursuant to EN 61 010 1 Overvoltage category II Degree of contamination 2 Permissible ambient temperature 5 to 40 C Permis...

Page 9: ...hing should be inserted into the ventilation system and the fan should not be blocked Caution Magnetic field For the closure mechanism both the door and the body of the sound proof box contain neodymi...

Page 10: ...for use Additional accessories as ordered see delivery note 2 2 Installation assembly Caution The sound proof box is too heavy to be set up by a single person The sound proof box should always be unpa...

Page 11: ...tton Place the perforated plate on the bottom of the sound proof box and slide the drip tray below 3 Operation The lighting and ventilation system of the sound proof box are switched on o using the pu...

Page 12: ...damage to the sound proof box 4 2 General use Open the door of the sound proof box completely Place the perforated plate on the bottom of the sound proof box and slide the drip tray below Position the...

Page 13: ...the sample vessels on the perforated plate or the supporting table AT 40 of the stand HG 40 If required use the non slip mat or similar 5 Cleaning and maintenance Cleaning and maintenance should be p...

Page 14: ...fective Fan defective Replacement required send in for repair if necessary Reduced noise protection Door not closed correctly Close door correctly clean magnetic closure Door seal damaged Replace door...

Page 15: ...r website as a PDF file www bandelin com Downloads 7 Accessories Correct accessories facilitate the application No obligation telephone consultation Website 49 30 768 80 0 www bandelin com A suitable...

Page 16: ...ontaminated before disposal Afterwards the device can be disposed of as electronic waste If decontamination cannot be carried out completely or correctly a safety data sheet of the substances used mus...