background image

1686-001 en/2022-05

Operating Manual

SONOREX DIGITEC

SONOREX DIGITEC

-RC

High-performance ultrasonic baths

Applies to:

DT 31, DT 31 H, DT 52, DT 52 H
DT 100, DT 100 H, DT 102 H, DT 102 H-RC, DT 103 H, DT 106
DT 156, DT 156 BH
DT 255, DT 255 H, DT 255 H-RC
DT 510, DT 510 H, DT 510 H-RC, DT 512 H
DT 514, DT 514 H, DT 514 BH, DT 514 BH-RC
DT 1028, DT 1028 H, DT 1028 CH
DT 1050 CH

Dokument: 1686-001

Sprache: en

Version: 2022-05

Summary of Contents for DT 31 H

Page 1: ...ths Applies to DT 31 DT 31 H DT 52 DT 52 H DT 100 DT 100 H DT 102 H DT 102 H RC DT 103 H DT 106 DT 156 DT 156 BH DT 255 DT 255 H DT 255 H RC DT 510 DT 510 H DT 510 H RC DT 512 H DT 514 DT 514 H DT 514...

Page 2: ...2 52 1686 001 en 2022 05 2022 1 2 GmbH Co KG Heinrichstr 3 4 12207 Berlin Germany Phone 49 30 76880 0 Fax 49 30 7734699 info bandelin com Certified in accordance with ISO 9001 and ISO 13485...

Page 3: ...perature 10 2 8 Danger due to ultrasound 10 2 9 Danger due to preparations used 11 2 10 Disposing of sonication fluid 11 2 11 Erosion of the oscillating tank 12 2 12 Preventing damage to the ultrasoun...

Page 4: ...egassing the sonication fluid DEGAS 26 5 9 Adding goods to be treated 26 5 10 Remove treated objects 27 5 11 Empty the oscillating tank 28 5 12 Enabling and blocking continuous operation 29 5 13 Troub...

Page 5: ...ese instructions for use before using the ultrasound bath Pay special attention to chapter 2 Safety If you pass this ultrasound bath on to others please enclose these instructions for use Contact your...

Page 6: ...asket or another suitable container An overview of suitable accessories can be found in chapter 10 Accessories In the case of stains discolouration rust etc a deep cleaning can be performed using spec...

Page 7: ...ltrasonic cleaning The ultrasound bath is not suitable for cleaning or disinfecting contact lenses The direct sonication of flammable liquids is not permitted Possible side effects restrictions Ultras...

Page 8: ...vice or protective equipment 2 4 Keep out of reach of children Children cannot identify hazards posed by the ultrasound bath Therefore keep the ultrasound bath out of the reach of children 2 5 Danger...

Page 9: ...connection at any time and without difficulty 2 6 Harmful to health due to ultrasound noise The ultrasound noise typical of the process can be perceived as very unpleasant Remaining within a radius of...

Page 10: ...tank which can be very hot After operation allow the ultrasound bath to cool down before lifting it for emp tying Non aqueous fluids can heat up many times faster than water A possible flashpoint can...

Page 11: ...e Observe the information on the label and in the safety data sheet of the preparation Keep the preparations away from children and untrained per sons 2 10 Disposing of sonication fluid Dispose of the...

Page 12: ...ishwashing detergents Chloride ions will corrode stainless steel Only use the ultrasound bath with accessories that are suitable for the device and the objects to be treated e g a basket Do not place...

Page 13: ...rence occurs with the operation of a wireless device move it further away from the ultrasound bath The ultrasound bath meets the requirements for class B devices according to EN 55011 2 14 Safety stic...

Page 14: ...2022 05 3 Design and function 3 1 Structure Fig 1 Overview of the device 1 Handles depending on model 2 Oscillating tank 3 Control panel 4 Filling level mark 5 Drain with ball valve depending on model...

Page 15: ...rside of the oscillating tank The ultrasound generates strong pressure fluctuations in the sonication fluid Cavitation bubbles are formed at the pressure minima At higher ambient pressure around the b...

Page 16: ...der no 3090 Various additional functions allow easy operation using a PC to control processes and for quality control The WINSONIC DT soft ware required for this is not part of the standard scope of d...

Page 17: ...t and work content see chapter 8 1 Technical data Adequate ventilation must be ensured The air supply under the bottom of the unit may not be obstructed If a sound proof box is used it must also ensur...

Page 18: ...off the ultrasound Result A noticeable noise can be heard when the ultrasound is switched on Contact our service department if you cannot hear any noise A foil test should be performed before the firs...

Page 19: ...sonicated are introduced in the contact liquid in an inset beaker for transfer of the ultrasound in the oscil lating tank The contact liquid in the oscillating tank must contain a surfactant preparati...

Page 20: ...ated Excessively long sonication can damage the surface of the goods to be treated Select a short sonication time where possible The ideal sonication time depends on several factors Type and concentra...

Page 21: ...dity condensation forms inside the device if you fill it with cold water Do not pour cold water into the ultrasonic oscillating tank when humidity is high NOTICE If you use an agent in powder form do...

Page 22: ...with a drain The device must be switched off Procedure 1 Fill the ultrasonic oscillating tank with water up to 1 3 2 Pour the correct quantity of agent into the ultrasonic oscillating tank See chapter...

Page 23: ...n time 5 To switch off the sonication press the start stop button Information The ultrasound switches off automatically as soon as the set sonication time has elapsed Then only the LED with the most r...

Page 24: ...ng of the sonication fluid is ideal See chapter 5 8 Degassing the sonication fluid DEGAS Before pre heating remove the basket or other accessories from the ultrasonic oscillating tank Cover the ultras...

Page 25: ...ctivated automat ically for 3 seconds every minute to mix the fluid to avoid retarda tion of boiling This function cannot be switched off You can configure the system so that this function is always a...

Page 26: ...e before sonication check whether the sonication fluid is contaminated If there is visible contamination replace the sonication fluid The sonication fluid must be degassed See chapter 5 8 Degas sing t...

Page 27: ...ects to be treated the surface of the ultra sound bath and accessories may be very hot Do not touch the surface of the ultrasound bath or accessories such as the lid Do not reach into the sonication f...

Page 28: ...illating tank if the soni cation fluid is visibly contaminated Also observe the specifications of the manufacturer of the prepa ration regarding the service life of the sonication fluid Replace used s...

Page 29: ...ous operation Requirement The mains plug is disconnected Procedure Press and hold down the button for setting the sonication time At the same time insert the mains plug into an earthed mains socket Th...

Page 30: ...of objects to be treated Uneven noises wobbling Inadequate fill level in the oscillating tank Slightly change the fill level of the sonication fluid in the oscillating tank In doing so observe the min...

Page 31: ...ions Rinse the oscillating tank thoroughly with water after each use Wipe dry with a soft cloth Clean edges and remove residues with a stainless steel care product without abrasive additives Do not us...

Page 32: ...e output of the ultrasound and the heating The output can be checked with a wattmeter between the device mains plug and the socket Procedure 1 Fill the ultrasonic oscillating tank with water 2 Switch...

Page 33: ...the filling level mark Temperature of the sonication fluid Degassing time Frame positioning Foil type brand thickness Sonication time Type and concentration of the ultrasound preparation Fluid for the...

Page 34: ...pecified by the manufacturer 2 Degas the sonication fluid See chapter 5 8 Degassing the sonication fluid DEGAS 3 Stretch aluminium foil house hold foil 10 m to 25 m thick over the foil test frame Depe...

Page 35: ...ed 9 Rinse the oscillating tank thoroughly to remove any detached foil particles You can order suitable foil test frames from BANDELIN electronic GmbH Co KG The foil test frames are configured for a w...

Page 36: ...nd bath if it has come into contact with toxic corrosive radioactive or biologically hazardous substances Also clean and decontaminate any accessories that you send in Download the Certificate of Deco...

Page 37: ...ly as elec tronic waste if it can no longer be used Do not dis pose of the ultrasonic bath with household waste Observe the locally applicable regulations for the disposal of electronic waste The vibr...

Page 38: ...Ultrasonic frequency 35 kHz Measuring accuracy of temperature sen sor 2 5 C Type Ultrasonic peak power ultra sonic nominal output Heating out put Power con sumption 230 V Power con sumption 115 V W W...

Page 39: ...5 DT 510 H RC 640 160 400 2 5 DT 512 H 860 215 400 2 7 5 4 DT 514 860 215 1 0 1 9 DT 514 H 860 215 600 3 6 7 1 DT 514 BH 860 215 600 3 6 7 1 DT 514 BH RC 860 215 600 3 6 7 1 DT 1028 1200 300 1 4 DT 1...

Page 40: ...5 DT 100 240 140 100 2 0 3 2 DT 100 H 240 140 100 2 0 3 3 DT 102 H 240 140 100 2 0 G 4 0 DT 102 H RC 240 140 100 2 0 G 4 0 DT 103 H 240 140 150 2 5 G 4 2 DT 106 240 130 4 0 G 5 2 DT 156 500 140 100 4...

Page 41: ...CH 500 300 300 30 0 G 23 3 DT 1050 CH 600 500 300 60 0 G 37 0 Type Internal dimen sions of oscil lating tank L W H Operating volume Connection for ball valve drain Weight mm l kg Overvoltage category...

Page 42: ...agent l 1 2 3 5 10 DT 31 DT 31 H 0 6 590 ml 10 ml 585 ml 15 ml 580 ml 40 ml 570 ml 30 ml 540 ml 60 ml DT 52 DT 52 H 1 2 1 1 l 15 ml 1 1 l 25 ml 1 1 l 40 ml 1 1 l 60 ml 1 0 l 120 ml DT 100 DT 100 H DT...

Page 43: ...0 ml 8 7 l 270 ml 8 5 l 450 ml 8 1 l 900 ml DT 514 BH DT 514 BH RC 12 5 12 3 l 130 ml 12 2 l 250 ml 12 1 l 380 ml 11 8 l 630 ml 11 2 l 1 3 l DT 1028 DT 1028 H 19 0 18 8 l 190 ml 18 6 l 380 ml 18 4 l 5...

Page 44: ...tects against contamination from outside Condensation water is guided into the ultrasonic oscillating tank Reduces noise Inset basket K EM made of stainless steel An alternative to DIN insert trays in...

Page 45: ...3 DT 103 H K 3 CL D 100 D 3 T KW 3 DT 106 K 6 D 6 DT 156 K 6 L D 156 DT 156 BH K 6 BL D 156 DT 255 H H RC K 5 C D 255 D 5 T KW 5 DT 510 H H RC K 10 GH 10 D 510 D 10 T KW 10 0 DT 512 H K 10 B D 510 D...

Page 46: ...timum utilisation of ultra sonic energy Beaker holder ES made of stainless steel For holding 4 inset beakers in larger ultrasonic baths Positioning for optimum utilisation of ultrasonic energy Impress...

Page 47: ...00 PK 3 C DT 106 KD 0 PD 04 SD 06 SD 09 PD 06 EB 05 DE 6 DT 156 KD 0 PD 04 SD 06 SD 09 PD 06 EB 05 DE 156 DT 156 BH KD 0 PD 04 SD 06 SD 09 PD 06 EB 05 DE 156 DT 255 H H RC KD 0 PD 04 SD 06 SD 09 PD 06...

Page 48: ...16 mm diameter Can also be used as a test tube stand Content of test tubes remains visible Tableting punch holder TH made of stainless steel For holding tableting punches with different diameters Sie...

Page 49: ...SH 7 DT 156 EK 10 EK 25 EK 50 EK 100 EK 250 GV 3 SM 6 FE 12 DT 156 BH EK 10 EK 25 EK 50 EK 100 EK 250 GV 3 SM 6 FE 12 DT 255 H H RC EK 10 EK 25 EK 50 EK 100 EK 250 GV 3 SM 5 FE 12 DT 510 H H RC EK 10...

Page 50: ...10 Accessories 50 52 1686 001 en 2022 05...

Page 51: ...1686 001 en 2022 05 51 52...

Page 52: ...1686 001 en 2022 05 1 2 GmbH Co KG Heinrichstr 3 4 12207 Berlin Germany Phone 49 30 76880 0 Fax 49 30 7734699 info bandelin com www bandelin com...

Reviews: