W
IRING
INSTRUCTIONS
FOR
BAMONIV TPS
OR
NIVAPRESS SGE 25
TO
CONNECTION
HEAD
BJSC
4. Tighten the sensor cable until the cable tie rests in the connection head
BAMONIV TPS
NIVAPRESS SGE 25
Tighten the cable gland until it seals
Tighten the cable gland until it seals
5. Reinstall the circuit board and tighten the screws
BAMONIV TPS
NIVAPRESS SGE 25
6. Apply cores according to the terminal assignment
BAMONIV TPS
NIVAPRESS SGE 25
Clamp 1: white (+)
Clamp 2: brown (-)
Clamp 3: green/yellow (shield)
Clamp 1: white (+)
Clamp 2: brown (-)
Clamp 3: green/yellow (shield)
7. Stow the ventilation hose in the connection head
BAMONIV TPS
NIVAPRESS SGE 25
ATTENTION: Do not kink hose!
Step not necessary
8. Insert cable for your process via the cable gland.
BAMO Mesures SAS ▪ 22, Rue de la Voie des Bans ▪ 95100 ARGENTEUIL
Tel. +33 (0)1 30 25 83 32 ▪ Fax: +33 (0)1 34 10 16 05 ▪ www.bamo.eu ▪ info@bamo.eu
590 msa590-04 SU3733-1.doc 07/18