Ignition Class "n"
11
Online-setting mode
CAUTION!
In this procedure the machine with the BTL system remains
operational, i.e., any change in the BTL output signal may
cause the machine to respond.
Be sure that this response will not result in any risk to persons
or equipment!
Maximum adjusting range for each adjustment procedure:
Start value = max. ±12.5 % of actual stroke,
End value = max. ±12.5 % of actual output value.
Note:
Between each adjustment procedure – also between setting of start
and end value – has to be exited by waiting 15 s (time out) until the next
adjustment can be carried out.
When the desired value cannot be reached within the first adjustment pro-
cedure because the adjustment range of ±12.5 % is exceeded, another ad-
justment can be carried out after 15 s. This can be repeated until the de-
sired value is achieved.
The BTL output signal at a particu-
lar magnet position is set to the
desired value which is then stored
by the controller as a start or end
value without having to power down
the entire system. For this reason
the safety advisory at right should
be especially noted.
Setting start value online:
1. Position system so that the mag-
net ring is located at the start posi-
tion.
2. Activate buttons:
Hold button 1 down for at least 3 s
and then – without releasing but-
ton 1 – hold both buttons down for
at least 3 s.
3. Setting the value:
Briefly pressing the buttons in-
creases or decreases the actual
value by approx. 1 mV or 2 µA.
Holding a button down for longer
than 1 s increases the step size.
4. Once the adjustment range or the
desired start value is reached, the
setup mode is automatically exited
if no button is pressed for at least
15 s. The buttons are again dis-
abled. Another adjustment proce-
dure can be carried out.
Setting end value online:
1. Position system so that the mag-
net ring is located at the end posi-
tion.
2. Activate buttons:
Hold button 2 down for at least 3 s
and then - without releasing but-
ton 2 - hold both buttons down for
at least 3 s.
3. Setting the value:
Briefly pressing the buttons in-
creases or decreases the actual
value by approx. 1 mV or 2 µA.
Holding a button down for longer
than 1 s increases the step size.
4. Once the adjustment range or the
desired end value is reached, the
setup mode is automatically exited
if no button is pressed for at least
15 s. The buttons are again dis-
abled. Another adjustment proce-
dure can be carried out.
Activate
buttons
current
position
value
new end
value
current
position
value
BTL5-... with magnet at end of stroke
The values shown here are examples only.
Displayed values (example)
at 0...10 V
at 4...20 mA
Activate
buttons
BTL5-... with magnet in start position
current
position
value
new start
value
current
position
value
The values shown here are examples only.
Displayed values (example)
at 0...10 V
at 4...20 mA