background image

– 25 –

If the pressure reducers are showing less than 50 bar o

2

 

or 30 bar n

2

o (with 10-litre cylinders), it is recommend-

ed to replace the relevant gas cylinder before a sedation. 
Apply the pulse oximeter to the patient. Slowly open the 
gas cylinders to the left and switch on the nitrous oxide 
mixer. 5 l/min. of pure oxygen flow automatically. you 
can reduce or increase the breathing bag fill level using 
the flow controller on the right. Switch on the suction 
unit and place the nasal mask on the patient once the 
bag is well inflated.
Keep  an  eye  on  the  breathing  bag. The  breathing  bag 
serves as a reservoir for the mixed gas and helps you 
monitor breathing and breathing capacity. It shows 
whether the gas flow rate is optimally adjusted in l/
min. the patient automatically takes in the mixed gas 
through inhaling. the breathing bag compresses each 
time the patient takes a breath. It should not be inflated 
too much, as then gas would escape and too much gas 
would be consumed. At the same time, the breathing 
bag should never deflate fully. If this happens, howev-
er, the valve for room air suction opens, ambient air is 
drawn in and the sedation cannot take effect properly. 
the folds at the side of the breathing bag should always 
be clearly visible. If the breathing bag is too full, reduce 
the total flow. If the bag is empty, increase the total flow 
immediately. 

During sedation, ensure the patient’s 

respiration is visible from the bag. 

If the bag does not 

show breathing movements, this may indicate that the 
mask is not positioned correctly and it should be reposi-
tioned. A small bag (e.g. 2.3 litres) is recommended when 
treating children. When treating adults, use a 3-litre bag.
you can now use the left concentration controller to 
add nitrous oxide to the mix, as learned in the advanced 
training course. titrate the nitrous oxide individually for 
each patient. to end the nitrous oxide sedation, press 
the “recovery” button and switch off the unit after a 

suf-

ficient recovery phase

 with pure oxygen. the length of 

the recovery phase depends on the length of sedation 
and maximum nitrous oxide concentration. the default 
is 5 minutes.

Please note:

It is not possible to make a general statement on how 
much nitrous oxide might be needed to sedate a patient. 
the necessary mixing ratio CAnnot be estimated on the 
basis of height and weight either. the greater a patient’s 
lung volume, the more nitrous oxide is required. Athletes 
usually need more nitrous oxide than non-athletes. the 

nitrous oxide concentration at which a patient is sedated 
may vary from day to day.

oUr tIP: Check how much nitrous oxide the patient 
received during his last sedation before any new 
sedation. you can use this as a rough guide. But you 
should still always start with a low dose.

hygiene measures:

oxygen-nitrous oxide mixer and Bag t

Use clean gloves during cleaning and ensure the envi-
ronment is clean and as free of dust as possible.
Clean the outsides of the mixer and the Bag t with a mild 
disinfectant approved for cleaning healthcare equip-
ment (no scouring agents). Do not spray the disinfectant 
directly onto the components to be cleaned. Instead, 
spray it onto a lint-free cleaning cloth which you can 
then use to wipe down the components. the cleaning 
agent must be wiped off at the end so that no residue is 
left. In case of the variants designed for installation only 
clean the display, frame and Bag t as described above.

other hygiene measures:

double nasal mask scavenger system  
and breathing bag:

Consult the relevant instruction manual for cleaning the 
scavenger system and bag.

7. hANDLINg AND hYgIENE

Summary of Contents for Touch 50

Page 1: ...0483 Made in Germany Instructions for use 02 N2 0 Mixer for Analgesia Baldus Touch 50 60 70 Baldus Bag T...

Page 2: ...sedation and medical gases Welcome to Baldus Medizintechnik GmbH Baldus Medizintechnik GmbH Auf dem Schafstall 5 56182 Urbar Germany 49 0 261 96 38 926 0 49 0 261 96 38 926 21 office baldus medizintec...

Page 3: ...g T 11 3 5 Back and side views 11 3 6 Technical specifications 13 4 Scope of delivery and accessories 14 5 Assembly instructions 15 6 Safety and regular controls 18 6 1 General safety instructions 18...

Page 4: ...ed 01 06 2017 0483 Inhalt nicht steril Contents not sterile Physicians using nitrous oxide have given our service a 1 1 4 rating You can trust the specialist in medical gases and inhalation sedation s...

Page 5: ...and con sult your instruction manual and or contact your ser vice partner or the manufacturer Baldus Medizintechnik GmbH 0049 261 96 38 926 0 Do not attempt to per form any repairs yourself or adjust...

Page 6: ...are facilities Take the table under Section 6 on reduc ing ambient air pollution through nitrous oxide into ac count A suction unit is compulsory and should have a suction power of at least 45 l min I...

Page 7: ...effects The general side effects of nitrous oxide are very minor e g vitamin B12 deficiency In case of continuous admin istration over two days Cardiovascular side effects No relevant side effects fr...

Page 8: ...te When creating a medical history please follow the content of the advanced training course Oxygen nitrous oxide mixer Controls and regulates the flow gas flow rate and per centage nitrous gas and ox...

Page 9: ...switched at any time 3 2 Main menu 1 Start sedation with patient ID A menu with letters and numbers appears for adding a patient by name or patient number After clicking Enter the sedation starts imm...

Page 10: ...last 20 error messages of the unit can be viewed and saved The memory can also be deleted All errors are displayed in the maintenance menu 4 7 Recovery time The recovery time can be set here individu...

Page 11: ...eathe in pure oxygen using the full face mask The demand valve of the mask opens on inhalation If a patient is unconscious and no longer breathing independently he can be actively ventilated by pressi...

Page 12: ...ormat FAT32 Possible sizes of up to 64 GB Fig 7 Baldus Touch view of the left side Fig 8 The full Baldus Touch system in the Baldus All in One Cart BACK FRONT All in One Cart N2 O pressure hose O2 pre...

Page 13: ...ation 0 Maximum N2 O concentration 60 Product variant 3 Baldus Touch 50 for installation Minimum O2 concentration 50 Maximum O2 concentration 100 Minimum N2 O concentration 0 Maximum N2 O concentratio...

Page 14: ...cer 16038 1 O2 Pressure Hose 16038 2 N2 O Pressure Hose 15015 3L or 15015 2 3L 3 L Breathing Bag for Adults or 2 3 L Breathing Bag for Children 15016 Baldus Holder for Cylinders e g cylinder trolley 1...

Page 15: ...keep a spare oxygen cylinder on hand For a centralised gas supply the gas cylinders and equip ment are located in a separate room and the gas flows to the treatment rooms through medical copper pipes...

Page 16: ...e used for this As the connectors are different sizes there is no possibility of confusion Assembling the Baldus Touch mixer and Baldus Bag T The Baldus Touch oxygen nitrous oxide mixer is in stalled...

Page 17: ...here 2 2 Now go from the point for the nozzle out 15 mm to the right and 15 mm upwards and draw the second point for the screws there 2 3 Now go from the point for the nozzle out 15 mm to the right an...

Page 18: ...ns and the Or dinance on Hazardous Substances Gefahrstoffverord nung The gas types must be checked after installing the mixer and accessories to rule out any mix up in the gas types To check the gas t...

Page 19: ...tage administer 100 oxygen to your patient to discontinue the nitrous oxide sedation and start the recovery phase If you do not have any ox ygen supply due to the power outage remove the mask from the...

Page 20: ...ow or there is a measure ment section failure The visual Check O2 signal appears in the top right of the sedation display and the O2 bar on the bottom right is shaded red A Check O2 gas supply pop up...

Page 21: ...an cause malfunctions Have annual safety checks and maintenance per formed including filter checks Service staff employees or partners of Baldus Mediz intechnik GmbH can use a service USB stick to acc...

Page 22: ...Leakscanbeeasilycheckedforusingaspecialleakdetec tion spray which is sprayed on all screw joints and con nections on the mixer and pressure reducers The leak detection spray consists of soap and wate...

Page 23: ...rs 5 Now close the oxygen cylinder 6 The Check O2 information signal appears after a short period of time 7 The system is vented Both pressure reducers now show 0 bar 8 You can now switch off the unit...

Page 24: ...n der is changed and at least monthly Close the gas cylinders completely there should be no appreciable drop in pressure for an hour Always open the high pressure valves of the gas cylinder s slowly O...

Page 25: ...the left concentration controller to add nitrous oxide to the mix as learned in the advanced training course Titrate the nitrous oxide individually for each patient To end the nitrous oxide sedation p...

Page 26: ...ople used to be administered a 100 dose of nitrous oxide at carnivals for fun after which they would stagger around the place comically This would make onlookers laugh j Can I sedate a patient with a...

Page 27: ...lable allen anwendbaren Anforderungen der Richtlinie 93 42 EWG Anhang II entspricht meets all applicable requirements of the Directive 93 42 EEC Annex II Name Adresse und Kennnummer mdc medical device...

Page 28: ...Baldus Medizintechnik GmbH Auf dem Schafstall 5 56182 Urbar Germany 49 0 261 96 38 926 0 49 0 261 96 38 926 21 office baldus medizintechnik de www baldus medical com...

Reviews: