– 20 –

2. Check n

2

o:

 If no nitrous oxide is being supplied, 

the unit emits visual and acoustic information signals. 
this also happens if the pressure is too low or there is a 
measurement section failure. the visual “Check n

2

o” sig-

nal appears in the top right of the sedation display and 
the n

2

o bar on the bottom right is shaded red. A “Check 

n

2

o gas supply” pop-up window appears and must be 

acknowledged. In addition, an acoustic information sig-
nal is audible. this event is recorded in the error memory 
log. Pure oxygen can still be administered to the patient, 
however the sedative effect will wear off without the ni-
trous oxide. the settings in the Sedation menu can still 
be changed. the alert disappears once n

2

o flows across 

the measurement section again. the check n

2

o informa-

tion signal of the unit is muted using the mute button. If 
the cause is an empty nitrous oxide cylinder, replace the 
cylinder. If the alert has been caused by something else, 
please contact the Service team.

6.3 Information signals and error codes

Information signals

the information signals are shown in the top right of the 
display.

1.

 

check

 

o

2

:

 

If no oxygen is being supplied, the unit 

emits 

visual and acoustic

 information signals. this also 

happens if the pressure is too low or there is a measure-
ment section failure. the visual “Check o

2

” signal appears 

in the top right of the sedation display and the o

2

 bar on 

the bottom right is shaded red. A “Check o

2

 gas supply” 

pop-up window appears and must be acknowledged. In 
addition, an acoustic information signal is audible. this 
event is recorded in the error memory log. the sedation 
continues, but without n

2

o and therefore without gas 

flow. Accordingly, the unit has a nitrous oxide lock: the 
unit switches off the nitrous oxide flow immediately if 
there is a lack of o

2

. the patient will not suffocate if the 

oxygen cylinder is empty and the practitioner forgets to 
remove the mask. Instead, the emergency valve on the 
Bag t opens and the patient breathes in the ambient 
air. the alert disappears once o

2

 flows across the meas-

urement section again. the check o

2

 information signal 

cannot be muted. the o

2

 signal sounds in greater inter-

vals when the mute button is pressed. If the cause is an 
empty oxygen cylinder, replace the cylinder. If the alert 
has been caused by something else, please contact the 
Service team. 

Fig. 19: Sedation display, check o

2

Fig. 20: Sedation display, check n

2

o

3. o

2

 mi

n

imum flow 1l/min.: 

Visual and acoustic sig-

nals: the o

2

 minimum flow is 1 l/min. the mixture ad-

ministered can never contain less than 1 litre of o

2

. this 

is displayed visually in the top right. the total mixture 
volume is never less than 3 l/min.

Fig. 21: Sedation display, o

2

 minimum flow 1l/min.

Summary of Contents for Touch 50

Page 1: ...0483 Made in Germany Instructions for use 02 N2 0 Mixer for Analgesia Baldus Touch 50 60 70 Baldus Bag T...

Page 2: ...sedation and medical gases Welcome to Baldus Medizintechnik GmbH Baldus Medizintechnik GmbH Auf dem Schafstall 5 56182 Urbar Germany 49 0 261 96 38 926 0 49 0 261 96 38 926 21 office baldus medizintec...

Page 3: ...g T 11 3 5 Back and side views 11 3 6 Technical specifications 13 4 Scope of delivery and accessories 14 5 Assembly instructions 15 6 Safety and regular controls 18 6 1 General safety instructions 18...

Page 4: ...ed 01 06 2017 0483 Inhalt nicht steril Contents not sterile Physicians using nitrous oxide have given our service a 1 1 4 rating You can trust the specialist in medical gases and inhalation sedation s...

Page 5: ...and con sult your instruction manual and or contact your ser vice partner or the manufacturer Baldus Medizintechnik GmbH 0049 261 96 38 926 0 Do not attempt to per form any repairs yourself or adjust...

Page 6: ...are facilities Take the table under Section 6 on reduc ing ambient air pollution through nitrous oxide into ac count A suction unit is compulsory and should have a suction power of at least 45 l min I...

Page 7: ...effects The general side effects of nitrous oxide are very minor e g vitamin B12 deficiency In case of continuous admin istration over two days Cardiovascular side effects No relevant side effects fr...

Page 8: ...te When creating a medical history please follow the content of the advanced training course Oxygen nitrous oxide mixer Controls and regulates the flow gas flow rate and per centage nitrous gas and ox...

Page 9: ...switched at any time 3 2 Main menu 1 Start sedation with patient ID A menu with letters and numbers appears for adding a patient by name or patient number After clicking Enter the sedation starts imm...

Page 10: ...last 20 error messages of the unit can be viewed and saved The memory can also be deleted All errors are displayed in the maintenance menu 4 7 Recovery time The recovery time can be set here individu...

Page 11: ...eathe in pure oxygen using the full face mask The demand valve of the mask opens on inhalation If a patient is unconscious and no longer breathing independently he can be actively ventilated by pressi...

Page 12: ...ormat FAT32 Possible sizes of up to 64 GB Fig 7 Baldus Touch view of the left side Fig 8 The full Baldus Touch system in the Baldus All in One Cart BACK FRONT All in One Cart N2 O pressure hose O2 pre...

Page 13: ...ation 0 Maximum N2 O concentration 60 Product variant 3 Baldus Touch 50 for installation Minimum O2 concentration 50 Maximum O2 concentration 100 Minimum N2 O concentration 0 Maximum N2 O concentratio...

Page 14: ...cer 16038 1 O2 Pressure Hose 16038 2 N2 O Pressure Hose 15015 3L or 15015 2 3L 3 L Breathing Bag for Adults or 2 3 L Breathing Bag for Children 15016 Baldus Holder for Cylinders e g cylinder trolley 1...

Page 15: ...keep a spare oxygen cylinder on hand For a centralised gas supply the gas cylinders and equip ment are located in a separate room and the gas flows to the treatment rooms through medical copper pipes...

Page 16: ...e used for this As the connectors are different sizes there is no possibility of confusion Assembling the Baldus Touch mixer and Baldus Bag T The Baldus Touch oxygen nitrous oxide mixer is in stalled...

Page 17: ...here 2 2 Now go from the point for the nozzle out 15 mm to the right and 15 mm upwards and draw the second point for the screws there 2 3 Now go from the point for the nozzle out 15 mm to the right an...

Page 18: ...ns and the Or dinance on Hazardous Substances Gefahrstoffverord nung The gas types must be checked after installing the mixer and accessories to rule out any mix up in the gas types To check the gas t...

Page 19: ...tage administer 100 oxygen to your patient to discontinue the nitrous oxide sedation and start the recovery phase If you do not have any ox ygen supply due to the power outage remove the mask from the...

Page 20: ...ow or there is a measure ment section failure The visual Check O2 signal appears in the top right of the sedation display and the O2 bar on the bottom right is shaded red A Check O2 gas supply pop up...

Page 21: ...an cause malfunctions Have annual safety checks and maintenance per formed including filter checks Service staff employees or partners of Baldus Mediz intechnik GmbH can use a service USB stick to acc...

Page 22: ...Leakscanbeeasilycheckedforusingaspecialleakdetec tion spray which is sprayed on all screw joints and con nections on the mixer and pressure reducers The leak detection spray consists of soap and wate...

Page 23: ...rs 5 Now close the oxygen cylinder 6 The Check O2 information signal appears after a short period of time 7 The system is vented Both pressure reducers now show 0 bar 8 You can now switch off the unit...

Page 24: ...n der is changed and at least monthly Close the gas cylinders completely there should be no appreciable drop in pressure for an hour Always open the high pressure valves of the gas cylinder s slowly O...

Page 25: ...the left concentration controller to add nitrous oxide to the mix as learned in the advanced training course Titrate the nitrous oxide individually for each patient To end the nitrous oxide sedation p...

Page 26: ...ople used to be administered a 100 dose of nitrous oxide at carnivals for fun after which they would stagger around the place comically This would make onlookers laugh j Can I sedate a patient with a...

Page 27: ...lable allen anwendbaren Anforderungen der Richtlinie 93 42 EWG Anhang II entspricht meets all applicable requirements of the Directive 93 42 EEC Annex II Name Adresse und Kennnummer mdc medical device...

Page 28: ...Baldus Medizintechnik GmbH Auf dem Schafstall 5 56182 Urbar Germany 49 0 261 96 38 926 0 49 0 261 96 38 926 21 office baldus medizintechnik de www baldus medical com...

Reviews: