Page
48
of
55
•
Lower the floor tray
•
Remove the new sachets from their packaging.
•
Whilst lifting the floor tray at the front, slide the sachets into the sachet holder.
•
Lower the floor tray, ensuring that the floor tray leg has located in the floor tray locator
•
Close the glove ports using the glove port covers.
•
Reinstall the Ezee Sleeves.
•
Reconnect the Bugbox Ax to the mains power supply and switch the Bugbox Ax on.
•
Allow the Bugbox Ax to reach an anaerobic environment in the workstation chamber by
leaving the Bugbox Ax for a period of approximately 1 hour.
•
Check that the environment in the workstation is anoxic using the anaerobic indicator
strips
•
If the environment in the workstation chamber is anaerobic, samples/ cultures may be
reintroduced to the workstation chamber via the interlock.
Replacing the mains plug fuse
–
UK users only
To replace the mains plug fuse;
•
Switch off Bugbox Ax and disconnect from the mains power supply.
•
Remove the plug from the mains socket.
•
Using a small flat bladed screw driver, remove the fuse cover from the mains plug. Figure
31 shows the fuse removal;
Figure 31: Mains plug fuse removal
•
Replace the fuse with a BS 1362 13A fuse.
•
Replace the plug in the mains socket.
•
Reconnect Bugbox Ax to the mains power supply and switch on the Bugbox Ax.