TROUBLESHOOTING
The Workstation will not Switch on
Check the workstation is plugged in
Check the Mains socket is switched on
Check the Circuit Breaker has not tripped
Check the Workstation fuses have not blown
Check the plug fuses have not blown
If all the above fail to switch the workstation on contact your local distributor.
The Circuit Breaker Trips
Remove any plugs from the internal sockets
Reset the Circuit Breaker
o
If the power does not come on unplug the workstation and check the mains
fuses
o
If the circuit breaker trips again contact your local distributor
o
If the power comes back on plug in any internal equipment again and check
that this does not trip the breaker. If it does then contact the appropriate
representative for that equipment.
Low Gas Pressure
The main causes for the workstation to identify low input gas pressures:
The gas input pressure is set to low
o
Check the gas supply bottle pressure is set to the recommended levels
defined in the ‘Gas Supply Requirements’ section
The gas bottle has run out
o
Check the gas supply bottle to make sure that it is not empty
A gas connection between the bottle and the workstation has come undone
o
Inspect the gas supply line from the workstation to the bottle to check there
it has not be disconnected
If the gas pressure alarm is activated during the interlock purge, there may be an issue
with input pressure drop during the purge cycle. This can be caused by:
The gas bottle running very low so it does not have the back pressure to maintain
the required flow rate
o
Change the gas bottle and re-run the purge
Restrictions in the supply of gas at pressure to the workstation
o
Review any restrictions and / or components between the gas supply and
the workstation
If all the above fail to clear the alarm then contact your local distributor or Ruskinn.