The correct cleaning agents must be used to clean the workstation. The use of incorrect
cleaning agents will damage the workstation and invalidate the warranty. The following
cleaning agents are permitted:
Ethanol, laboratory grade at a maximum concentration of 70% by volume ethanol
in distilled water.
Isopropanol, laboratory grade at a maximum concentration of 70% by volume
Isopropanol in distilled water.
Tristel Fuse Sachet, 1 sachet diluted in 3 litres of distilled water, or Tristel Duo
Foamer. Tristel Fuse Sachets and Duo Foamer are available from Ruskinn
Technology Limited.
Ruskinn Technology Limited anti-static cleaner.
Distilled or de-ionised water.
No other cleaning agents are permitted.
The use of UV light is prohibited in the workstation as it will cause permanent damage.
The use of UV Light within the Workstation will invalidate the warranty.
Cleaning procedure – during and after each use
During use, clean any spills immediately using paper towels soaked in an appropriate
cleaning agent. Wipe dry using a dry paper towel.
After each use;
Remove all waste materials from the workstation chamber(s).
Wipe the workstation chamber floor tray(s) using paper towels soaked an
appropriate cleaning agent and squeezed to remove excess fluid.
Wipe the workstation chamber floor tray(s) clean using paper towels to dry.
Wipe the interlock floor tray using paper towels soaked in an appropriate cleaning
agent and squeezed to remove excess fluid.
Wipe the interlock floor tray using paper towels to dry.
If the workstation has a waste port then check the waste bag and replace if full.