Bugbox
UM-16 R5.0
Page 40 of 48
Bugbox User Manual
Replacing the mains fuses
To replace the mains fuses;
Switch off the Bugbox.
Remove the plug from the mains socket.
Remove the mains fuse drawer using a small flat bladed screwdriver. Figure 29 shows the
removal of the mains fuse holder;
Figure 29: Mains fuse holder removal
Replace the mains fuses. The fuse ratings are;
o
240V 50Hz – F3.15A H250V
o
220V 60Hz - F3.15A H250V
o
110V 60Hz – F5A H250V
o
100V 50/60Hz – F5A H250V
Fuses should be fast blow, for example Cooper Bussmann S501. For more information, contact your
local distributor.
Replace the mains fuse holder. The fuse holder will click when it is fully inserted.
Replace the plug in the mains socket.