PanaFlow Z3 User’s Manual
25
Chapter 3. Programming
3.5.2
Standard Analog Output
The PanaFlow Z3 has one
Analog Output
and one
Digital Output
in Standard configuration.
3.5.2.1
Setting up Analog Output
The PanaFlow Z3 meter has one Analog Output in standard configuration. For additional Analog outputs Optional I/O
boards may be purchased. Use steps as in section “Log-in and Primary Pages” to navigate to the
Input/Output
settings page.
1.
Select
[Analog Output (Std)]
and press
[ENTER]
.
2.
If you do not wish to connect an Analog Output, you should turn Analog Output selection to OFF.
3.
If you are connecting an Analog Output, choose 4-20mA option. The Figure 23 below, shows the available
options.
4.
Select the Measurement to be sent out on the 4-20mA output, followed by the
[Base Value]
and
[Full Value]
selection. Refer Table 2 to see measurement options available for Analog output.
5.
Select
[Error Handling]
. Refer to Section 3.5.2.2 to choose an option that suits your needs.
3.5.2.2
Understanding the Error Handling Option
The following Table 3 shows the response to each of the Error Handling options. For a multi-channel meter, the [Path
Error Handling] set to ON (see Figure 36) changes the Output response. See Table 4 for Analog Output response with
Path Error Handling set to ON.
Note:
Table 4 assumes Composite Actual Volumetric is chosen as Measurement for Analog Output.
Table 2: Measurement Options for Analog output
Measurement Channel
Measurement Options for Analog Output
Composite
Velocity, Actual Volumetric, Standard Volumetric, Mass Flow, Average
Volumetric Flow Rate, Sound speed, Reynolds#
Channel x
Velocity, Actual Volumetric, Standard Volumetric, Mass Flow, Average
Volumetric Flow rate, Sound speed, Standard Deviation of Gain, Gain and
SNR
Table 3: Analog Output Error Handling options
Option
Output Response
Low
Forces Output to 4mA on error
High
Forces Output to 20mA on error
Hold
Holds the last “good” reading
Other
Enables the user to enter a value between 4mA and 20mA, to be output
during an error
Table 4: Analog Output Error Handling with Path Error Handling set to ON
Ch1 in Error
Ch2 in Error
Ch3 in Error
Error displayed in
Meter
Expected
Composite
Volumetric(Act)
behavior
Analog Output Response
No
No
No
No Error
Measured
Composite
Volumetric(Act)
mA proportional to the
Measured Composite
Volumetric(Act)
Yes
No
No
E22:
SingleChAccuracy
Measured
Composite
Volumetric(Act)
mA proportional to the
Measured Composite
Volumetric(Act)
Summary of Contents for Panametrics PanaFlow Z3
Page 1: ...Flow 910 311 Rev C April 2019 Panametrics com PanaFlow Z3 User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 8: ...Contents vi PanaFlow Z3 User s Manual...
Page 12: ...Preface x PanaFlow Z3 User s Manual no content intended for this page...
Page 21: ...PanaFlow Z3 User s Manual 9 Chapter 2 Installation Figure 8 Lifting the PanaFlow Z3 System...
Page 25: ...PanaFlow Z3 User s Manual 13 Chapter 2 Installation Figure 12 Remote Mount Wiring Diagram...
Page 26: ...Chapter 2 Installation 14 PanaFlow Z3 User s Manual no content intended for this page...
Page 58: ...Chapter 3 Programming 46 PanaFlow Z3 User s Manual no content intended for this page...
Page 86: ...Appendix A Specifications and Model Configurations 74 PanaFlow Z3 User s Manual...
Page 104: ...Appendix B Data Records 92 PanaFlow Z3 User s Manual no content intended for this page...
Page 108: ...Appendix D CE Mark Compliance 96 PanaFlow Z3 User s Manual no content intended for this page...
Page 110: ...Warranty 98 PanaFlow Z3 User s Manual no content intended for this page...
Page 116: ......
Page 117: ......