PanaFlow Z3 User’s Manual
27
Chapter 3. Programming
Figure 23: Setting up Analog Output
3.5.2.3
Calibrating Analog Output
To calibrate Analog Output, use steps as in section “Log-in and Primary Pages” to navigate to the
Input/Output
settings page. You may use a multimeter or DCS/SCADA to calibrate the Analog Output. Regardless of whether
multimeter or DCS/SCADA is used, the steps below remain the same. For better readability, the steps below only
indicate multimeter and does not repeat multimeter or DCS/SCADA.
1.
Turn ON the multimeter (if used), and set it to measure Current (mA) DC. Connect test lead from positive side
(Aout+) of the Main Analog Output to the positive terminal of multimeter, and the negative lead to the negative
terminal (Aout-).
2.
In Meter Menu, select
[Analog Output (std)]
and press
[ENTER]
. Then select
[4-20mA]
and press
[ENTER]
.
3.
Scroll down and select
[Calibrate]
option.
4.
Select
[Calibrate 4mA]
and check if the reading on the multimeter reads 4.00mA ± 0.01mA. If the value on the
multimeter is not 4.00mA ± 0.01mA, input the value read on the multimeter into the
Calibrate 4mA
value and
press
[ENTER]
. Check multimeter again verify the Current reads 4.00mA within ±0.01mA.
5.
Select
[Calibrate 20mA]
and check if the reading on the multimeter reads 20.00mA ± 0.01mA. If the value on the
multimeter is not 20.00mA ± 0.01mA, input the value read on the multimeter into the
Calibrate 20mA
value and
press
[ENTER]
. Check multimeter again verify the Current reads 20.00mA within ±0.01mA.
Analog Output (Std)
Off
Measurement Type
Channel 1
Channel 2
Channel 3
Composite
Full Value
Full Value
Full Value
UNIT: ft/s
3.0000
3.0000 ft/s
Calibrate 4mA
UNIT: mA
4.000
Error Handling Value
Value
UNIT: mA
4.00
Calibrate
Calibrate 4mA
Calibrate 20mA
Percentage of Scale
Measurement Type/Comp
Measurement Type
4-20 mA
Calibrate
Base Value
Error Handling
Full Value
4-20 mA
Calibrate
Base Value
Error Handling
Full Value
4-20 mA
Calibrate
Base Value
Error Handling
Full Value
Input/Output
Analog Output (Std)
Digital Output (Std)
Option Comm (Slot 1)
Option I/O (Slot 2)
Error Handling
Low
High
Hold
Other
4-20mA
4-20mA
Base Value
Full Value
Calibrate
Measurement Type
Channel 2
Channel 3
Composite
Channel 1
Avg Vol Flow Rate
Soundspeed
Mass Flow
Measurement Type/Ch #1
Volumetric (Act)
Volumetric (Std)
Mass Flow
Velocity
Measurement Type/Ch #1
Soundspeed
Gain
Avg Vol Flow Rate
Std Deviation Gain
4.00 mA
Value
9
2
9
2
9
2
9
2
9
2
9
2
9
2
9
or
9
2
9
or
9
2
9
or
9
2
9
or
9
2
9
or
9
2
Volumetric (Act)
Volumetric (Std)
Mass Flow
Velocity
Measurement Type/Comp
Reynolds #
Measurement Type/Ch #1
Std Deviation Gain
SNR Up
Gain
SNR Dn
[ ]
2
9
[ ]
UNDO
MOVE
[ ]
[ ]
SAVE
ALTER
[ ]
2
9
[ ]
UNDO
MOVE
[ ]
[ ]
SAVE
ALTER
[ ]
2
9
[ ]
UNDO
MOVE
[ ]
[ ]
SAVE
ALTER
Operator Access or Higher
Summary of Contents for Panametrics PanaFlow Z3
Page 1: ...Flow 910 311 Rev C April 2019 Panametrics com PanaFlow Z3 User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 8: ...Contents vi PanaFlow Z3 User s Manual...
Page 12: ...Preface x PanaFlow Z3 User s Manual no content intended for this page...
Page 21: ...PanaFlow Z3 User s Manual 9 Chapter 2 Installation Figure 8 Lifting the PanaFlow Z3 System...
Page 25: ...PanaFlow Z3 User s Manual 13 Chapter 2 Installation Figure 12 Remote Mount Wiring Diagram...
Page 26: ...Chapter 2 Installation 14 PanaFlow Z3 User s Manual no content intended for this page...
Page 58: ...Chapter 3 Programming 46 PanaFlow Z3 User s Manual no content intended for this page...
Page 86: ...Appendix A Specifications and Model Configurations 74 PanaFlow Z3 User s Manual...
Page 104: ...Appendix B Data Records 92 PanaFlow Z3 User s Manual no content intended for this page...
Page 108: ...Appendix D CE Mark Compliance 96 PanaFlow Z3 User s Manual no content intended for this page...
Page 110: ...Warranty 98 PanaFlow Z3 User s Manual no content intended for this page...
Page 116: ......
Page 117: ......