Copyright 2013 Baker Hughes Company.
100 | DPI
620G Instruction Manual–English
Chapter
10.
FOUNDATION™ Fieldbus
4. The main FOUNDATION™ Fieldbus application will now open.
Note:
Updating CH1, CH2, P1 or P2 tasks is only possible when the FOUNDATION™
Fieldbus application is not running. Close the FOUNDATION™ Field bus application by
pressing the Home
button and selecting Exit. Refer to Section
Note:
Ensure the DPI
620 Genii USB setting is in Storage Device mode. See Section
Note:
CH2 is locked into Volts measure mode. Trying to select a function on CH2 will result
in an information message and the selected function not being set.
10.3 FOUNDATION™ Fieldbus Toolbar
On entering the FOUNDATION™ Fieldbus application, the toolbar is displayed. The icons are
grayed out when they are not active.
Figure
10
-
3: FOUNDATION™ Fieldbus Toolbar
The icons functions are described below:
10.4 Scanning for Devices
The following steps describe how to scan for FOUNDATION™ Fieldbus devices over a
FOUNDATION™ Fieldbus H1 connection:
Icon
Name
Description
OPEN CONNECTION
Only available when awaiting an open connection.
(When navigating devices Open Connection is
replaced by the Close
icon).
CLOSE
Only available in navigation tree (refer to Section
10.8)
and Functional Group view (refer to Section
10.9). This
icon closes down the connection and returns to Device
Focus view (refer to Section
10.7).
SETTINGS
Application configuration settings and DD library details
(refer to Section
10.14).
COMMIT
Commit updated values back to the device (refer to
ABORT
Abort Update of parameters, reverts to previous values
(refer to Section
10.9.3).
STATUS
Provides device profile of currently connected device
(refer to Section
FUNCTION FINDER
Search FF variables and device functions.
HOME
Returns to Main Application. Allows user to either
minimize or exit. If FOUNDATION™ Fieldbus readings
are to be referenced in Main Application, then minimize
should be selected.
DEVICE COMMUNICATIONS ON
Indicator showing active communication.
DEVICE COMMUNICATIONS OFF
Indicator showing no active communication.