Programming Page
29/44
MID_M2000_BA_02_1004
Advanced
Empty Pipe Cal.
Note: To compensate different fluid conductivity, signal cable lengths or
sizes during measurements, you can calibrate them accordingly. This is
important in case that fluid monitoring has been activated and “empty
pipe” is signalized although the pipe is filled.
Proceed as follows:
1. Select “Empty Pipe Cal“
2. Set calibration to “CAl [on]“
3. Observe the voltage “Volt“
4. When stable, select “Store“ and press
E
5. Fill pipe with fluid
6. Select “Cal. Pipe Full“
7. Set calibration to “Cal [AN]“
8. Observe the voltage “Volt“
9. When stable, select “Store“ and press
E
10.
Password
Security
There are three possible access levels, each with its own unique
password:
•
Administrartor
PIN
•
Service
PIN
•
User PIN
The password security consists of a five-digit PIN and is set at the
factory to [00000]. Enter a number above zero to activate password
security. Activate password security in the following order:
Administrator, Service, User.
Note: You cannot activate the user password before administrator and
service passwords have been activated.
About 5 minutes after configuration, password security becomes active.
As soon as password security has become active, a PIN is required for
configuration. Depending on the individual PINs, you are either in the
administrator, service or user levels with corresponding access rights
(identified in the operating manual by the lock A, S and B).