Programming Page
28/44
MID_M2000_BA_02_1004
Advanced
Factory
setting
The analog output was already set at the factory to
exactly 4.00 mA (zero) and 20.00 mA (margin). In
case that you wish to adapt your external system,
use the item “corrective factor“.
Software Filter
This filter is intended to help eliminate unrequested peaks during
measurement.
Activation
Activates or deactivates the software filter
Filter Delay
Filter Delay lets you set the amount of time that the
flow will be held constant once the filter is activated
(change of flow per time unit).
Acceleration
Factor
This parameter lets you set the maximum
acceleration for a given pipe diameter (change of
flow per time unit). If the realized fluid acceleration
exceeds the configured maxi-mum acceleration,
fluid flow will be held constant for the time set at the
Filter Delay parameter.
Constant Flow
This parameter lets you set the acceleration limit for
a constant flow.
Peak Detect
Peak Defect offers a diagnostic view of acceleration
components observed during flow condition and
records the “high water mark“. Press
+
button to
reset.