Cable connections
Page 5 / 23
MID_VD_BA_02_1607
M
1
0
0
0
-2
3
3.2 ModMAG
®
M1000
Communication settings
ModMAG
®
M1000 port adjustment
Navigate to "Main Menu" / "Communication" and adjust as follows:
•
Interface:
ModBus
®
RTU
•
Port address:
1
•
Mode:
RS232
•
Baud rate:
9600
•
Parity:
Even
Check if the hardware DIP switches adjusted for a RS232 interface.
Switch off the amplifier before connecting the cable harness of the Verification Device.
Opening the cover:
1. Using a 1/4" slotted screwdriver, remove the two upper screws from front of the
amplifier.
2. Loosen the two lower screws until the screw heads protrude above the surface of
the amplifier door.
3. Open the amplifier (lift up the door).
Connecting the cable harness:
The individual connector wires are labeled as to where each connector is to be
connected onto the internal circuit board of the amplifier. A connection instruction label
has been placed inside of the amplifier for connection reference.
On ModMAG
®
M2000 cable harness are following connectors:
1. Output 1 & 2 / Input (6 pins)
2. RS232 (5 pins)
3. Analog output (3 pins)
4. Detector electrode (5 pins)
5. Detector coil (3 pins)
6. Amplifier electrode (5 pins)
7. Amplifier coil (3 pins)
8. Detector ground (alligator clip)