Technical data
Page 22 / 23
MID_VD_BA_02_1607
9. Technical data
9.1 Features
The Field Verification Device
• determines if the ModMAG
®
M2000 or B-MAG I M5000 is within one percent of the
original factory calibration.
• verifies the functionality of all of the meter’s inputs and outputs.
• measures electrode resistance and integrity.
• measures coil resistance and integrity.
• measures coil insulation resistance.
• measures current and frequency at previously selected flow rate.
• evaluates the signal processing functionality.
• verifies that there is no high-voltage noise disturbing the coil circuit.
• provides certified printout showing test results.
• is capable to store 100 meter tests.
9.2 Accessories
• Test cable for ModMAG
®
M2000 connection
• Test cable for B-MAG I M5000 connection
• AC charger with EU/US connection adaptors
• Automobile utility DC power supply connector
• PC communication cables to Verification Device
• Pelican suitcase for transport
9.3 Accuracy
9.3.1 Amplifier
9.3.2 Detector
Input gain
±
0,1%
Excitation coil current
20-400mA;
±
0,1%
Excitation coil frequency
50/60Hz;
±
1%
Analogue outputs current
4-20mA;
±
0,1%
Digital outputs
1Hz/20 kHz; – 0,1%
Coil resistance
40-120 Ohm +/-1 %
Electrode resistance
(against ground)
100 kOhm-10 MOhm; +/-10 %
Coil insulation
(against ground)
100 kOhm-10 MOhm; +/-10 %