Programming
Page 21/43
UF_iS4000_BA_02_1811
Output 1 /2
Set Max
The Flow Max Set Point establishes as a percentage of full scale
flow, the max. threshold at which the output alarm will be activated.
You can freely select thresholds in 1% steps. Flow rates
below/above the threshold will activate the output alarm.
Flow
Simulation
Flow Simulation provides analogue and digital output simulation based on a
percentage of the full scale flow in cases where no real flow is occurring. The range
of simulation includes 0% to +100% in steps of 10% of the full scale flow. This
function still remains active once you have left the menu. It is necessary to set it to
“Off” to deactivate it. If the simulation is still active, a character “S” will be displayed
in the measuring mode.
Clear total
Total
The totalizer is reset within the “ClearTot” item or flow meter tool software.