Changing Factory Settings
Changing the Scale Factor
MPORTANT
I
Changing the Scale Factor changes the accuracy of the meter, potentially causing over-dispensing or under-dispensing of the liquid.
This has the potential to cause a mechanical breakdown.
Scale Factor Digits
1 Press
RESET
to advance through the Scale Factor digits
2 Press
TOTAL
to change the selected number
OTEE:
N
The range of possible Scale Factors is 0 0000…1 9999
3 Press
RESET
advance to the next number in Scale Factor
4 Repeat steps 2 and 3 for all five digits in Scale Factor
5 If no Scale Factor changes are necessary, press and hold
TOTAL
and
AUTO
until the
display flashes three times, then goes blank
OR
Press
RESET
to advance to the
EPM Manual Mode Enable
screen
Verifying Changes
Press and hold
TOTAL
and
AUTO
simultaneously to verify that the Scale Factor is correct
Enabling or Disabling the EPM Mode
To enable or disable
EPM Mode
:
1 Press
RESET
, the Auto icon displays
2 Press
AUTO
to turn the Auto icon on and off
When the icon is turned on the meter can be put in
EPM Mode
through the special key
sequence on the meter keypad When the Auto icon is turned off, the key sequence is
disabled and the meter cannot be put in
EPM Mode
3 Press
RESET
to advance to the
Network Program
screen
Unit of Measure
Scale Factor Digits