The Informant™ 2
6
0019-9211 Rev 9
NOTE:
For dual detectors, the red probe tip will not screw onto a
probe that has a refrigerant sensor installed. This feature
safeguards against the possibility of an operator mistakenly thinking
that a detector with a refrigerant sensor is configured to detect
combustibles. See the WARNING below.
2.3. Refrigerant Sensor Installation
Install the refrigerant sensor, filter, and “blue” probe tip onto the end of
the detector’s flexible probe as shown in Figure 3a. Observe the blue
color coding on both the sensor and probe tip.
WARNING:
Explosion hazard. DO NOT use the Informant 2 with
its
refrigerant
sensor installed in an area that is classified by the
National Electrical Code as being a hazardous location where a
potential for explosion or fire exist because of flammable gases or
vapors in the atmosphere.
COMBUSTIBLE REFRIGERANT GAS WARNING:
For your
safety, DO NOT use the Informant 2 with the refrigerant sensor to
detect refrigerants which are rated as combustible/flammable
gases (e.g. ASHRAE - A2 or A3 rated refrigerants). Use with
combustible/flammable refrigerant gases can cause an explosion
resulting in death or severe injury.
Figure 3a. Sensor Installation (Refrigerant Sensor)