background image

Model 5100E 

 Introduction and Overview 

 

1.3.6

 

Sensor 

The  Neutronics  Inc 

ZR-500E  mini  Zirconia

  sensor  for  measuring  percent  range  oxygen  is  a  cost  effective 

solution for many applications.  It is a solid-state device based on a Zirconia solid electrolyte sensor with a 
low  power,  on-board  heater,  and  two  electrodes.    Unlike  many  solid-state  oxygen  sensors,  the  ZR-500E 
does not require a reference gas for normal operation (see calibration requirements). 

The  sensor is mounted directly onto a printed circuit board that supplies bias and heater voltages to the 
sensor  element.    A  small  capillary  on  the  sensor  surface  controls  diffusion  of  oxygen  into  the  sensor.  At 
operational  temperature,  oxygen  is  electrochemically  reduced,  causing  current  flow  through  the  solid 
electrolyte.    Oxygen  concentration  in  the  sample  gas  is  determined  by  measuring  the  current  flowing 

through the sensor’s two electrodes.

 

H

H

S +

S-

A

Ammeter

(Oxygen Analyzer)

Sensor Bias

Voltage

Heater Power

 

 

The  robust  design  of  the  ZR-500E  sensor  assures  accurate  measurement  as  well  as  a  quick  response 
characteristic to serve a wide range of oxygen measurement applications.   Its expected operating life is 2 
years under normal operating conditions, and it has an unlimited shelf life.  It is not affected by position, 
and it may be exposed to high atmospheric pressures with no ill effect on performance. 

1.3.7

 

Sensor Flow-Through Head 

The sensor mounting-base allows the Model ZR-500E sensor to be used for process monitoring. It serves as 
both  the  receptacle  for  the  sensor  and  the  delivery  system  for  a  gas  sample  from  a  process  vessel  or 
stream.  It includes a sample inlet, a flow-through chamber a sample exhaust, and a screw-in receptacle for 
the oxygen sensor. 

1.3.7.1

 

Sample Gas Inlet 

A sample gas inlet  port is provided for installation with a process oxygen sampling system.  Gas must  be 
directed from the measured process to the sample inlet port via positive pressure from the source, or an 
external  pump.    The  Model  ZR-500E  sensor  and  flow-through  head  combination  can  be  installed  in  any 
Neutronics Inc. Process Sampling System. 

1.3.7.2

 

Sample Gas Exhaust 

A sample gas exhaust port is provided for installation with a process oxygen sampling system.  Gas must be 
directed from the sample exhaust port to an exhaust at ambient pressure.  Any change in pressure to the 
exhaust port of the analyzer will affect its reading and accuracy. 

Figure 3 

ZR-500E sensor circuit 

Summary of Contents for Neutronics 5100E

Page 1: ...www neutronicsinc com High Purity Instruments MODEL 5100E OXYGEN ANALYZER CONTROLLER PERCENT RANGE OPERATIONS MANUAL File Name MN A 0008 Manual Part Number C5 06 4900 08 0 Revision Level C Revision D...

Page 2: ......

Page 3: ...4 7 Analog Current Output 1 9 1 4 8 Range ID Output 1 9 1 4 9 Service Port 1 9 1 5 CONTROL PANEL USER INTERFACE 1 9 1 5 1 The UP Pushbutton 1 9 1 5 2 The DOWN Pushbutton 1 9 1 5 3 The MODE Pushbutton...

Page 4: ...TING 4 1 4 1 SYSTEM SETUP 4 1 4 1 1 System Setup via Control panel Keypad 4 1 4 1 2 System Setup via Service Port 4 3 4 1 3 Change factory settings via Hardware Jumpers 4 16 4 2 ROUTINE PERIODIC MAINT...

Page 5: ...l down before attempting to access the sensor Ensure the Model 5100E has been properly calibrated before use Never expose the Model 5100E analyzer chassis to water high humidity or moisture The analyz...

Page 6: ...ave questions or comments we would like to hear from you Neutronics Inc Customer Service Department 456 Creamery Way Exton PA 19341 Tel 610 524 8800 Toll Free 800 378 2287 US only Fax 610 524 8807 EMA...

Page 7: ...sensor to be installed close to a sampling point for the fastest response time possible in process monitoring and control applications A flow through mounting head is supplied for use with all Neutron...

Page 8: ...ERVICE BY AUTHORIZED TECHNICIAN ONLY RISK OF ELECTRICAL NEMA 4 Panel mounting Gasket Removeable Terminal Blocks Remote Sensor Module Interface Mounting Studs Rear view Chassis Power Supply Board Relay...

Page 9: ...ly is available for installations where a DC voltage is required to power the Model 5100E A 12 VDC battery backup power input battery not provided is also provided to act as an emergency back up in ca...

Page 10: ...ement applications Its expected operating life is 2 years under normal operating conditions and it has an unlimited shelf life It is not affected by position and it may be exposed to high atmospheric...

Page 11: ...nt component of the electrostatic discharge ESD shielding design Since the Model 5100E is a flush mounted system the portion of the instrument housed in the chassis will be located behind the control...

Page 12: ...y Board RS 232 COM Line Mains Power Input Power Supply Main Board Display Board Control Panel Fault Relay Outputs Alarm 1 2 Relay Outputs Range IDOutput Analog Voltage Output 4 20 mA Output Remote Sen...

Page 13: ...owable or descending lowest oxygen level allowable action The relay output may be configured for failsafe relay coil de energized in alarm state or non failsafe relay coil energized in alarm state act...

Page 14: ...em setup Available settings are 0 1VDC 0 5VDC or 0 10VDC The factory default setting is 0 1 VDC The Analog voltage output is scaled according to the analyzer s selected range and must be used in conju...

Page 15: ...tion output The range ID output is used in conjunction with the Analog voltage and Analog current outputs when auto ranging is used It provides an indication of the Analog outputs selected full scale...

Page 16: ...of the sample gas and updating the display 1 5 6 Alarm 1 Indicator LED The purpose of the Alarm 1 Indicator LED is to inform the user via the control panel that the measured oxygen concentration has e...

Page 17: ...l C ECO 8029 File Name MN A 0008 Page 2 1 2CHAPTER 2 SYSTEM INSTALLATION AND START UP 2 1 Installing the Analyzer STEP 1 LOCATE THE ANALYZER PANEL CUTOUT STEP 2 INSTALL THE REMOTE SENSOR STEP 3 ELECTR...

Page 18: ...unit into the cutout rear chassis first and seat the control panel gasket on the mounting surface The gasket on the analyzer control panel ensures a watertight seal around the control panel cutout Sec...

Page 19: ...and exhaust fittings SENSORRECEPTACLE O RINGTOP SEAL 1 8 FNPT SAMPLE EXHAUST GASKET MOUNTINGSCREW INSERT 6 32 1 8 FNPT SAMPLEINLET THREADEDCONNECTOR 3 4 16 SENSOR MOUNTINGHOLES 2 PLACES 6 32 THREAD 36...

Page 20: ...r connecting the sample exhaust to vent use 1 8 or 1 4 rigid tubing and 1 8 MNPT fittings of a material compatible with process gas composition Select a vent location that is known to be at atmospheri...

Page 21: ...ver the 3 4 16 threaded connector Do not lubricate the o ring Verify the supplied gasket is seated in the bottom of the flow through head sensor receptacle Screw the sensor into the flow through head...

Page 22: ...certain that no flammable gases are present in the area where the Model 5100E analyzer will be installed CAUTION The Model 5100E housing is not rated waterproof Do not mount the analyzer or the sensor...

Page 23: ...arrangement is affixed to the top of the chassis for easy reference during installation and maintenance VAC configuration shown below The terminal blocks feature screwed terminals The terminal blocks...

Page 24: ...3 TB3 4 5 6 7 8 9 10 11 Bat BU V Out mA Out RS 232 TX RX RTN 12 VDC no connection no connection Service Port NO Normally Open NC Normally Closed C Common VACPOWERINPUT 90 264 VAC 47 63 Hz 10 Regulatio...

Page 25: ...ay Output Connections from the Fault relay contacts to the user s process control equipment are made at terminal block TB2 on the rear of the analyzer chassis The Fault relay contacts are voltage free...

Page 26: ...ain to match the terminal pins against the terminal ID label on the top of the analyzer chassis Use 20 AWG 2 conductor stranded wire twisted pairs for the connections It is not necessary to use shield...

Page 27: ...ANALYZER TB2 CONNECTION SIGNAL DESIGNATION AT COMPUTER COMPUTER DB9 SERIAL PORT CONNECTION TX Pin 9 RX Pin 2 RX Pin 10 TX Pin 3 RTN Pin 11 RTN Pin 5 Serial Cable Ferrite Core Cable Tie Cable Tie 2 1 3...

Page 28: ...tem away from exposure to rain dripping water or hose down Correctly installed all of the wiring Connected the sensor interface cable at both the sensor and analyzer Ensured gas tight plumbing at the...

Page 29: ...s and enter into the appropriate system mode according to programmed parameters refer to the Software Setup Log that is supplied by the factory for the configuration details of your Model 5100E Once...

Page 30: ...ontrol panel section 4 1 1 or the service port section 4 1 2 2 2 3 STEP 3 Set Alarm 1 and Alarm 2 After the unit has been calibrated on known gas sources set the alarm thresholds according to process...

Page 31: ...the control panel in the aforementioned order by repeatedly pressing and releasing the MODE key When a user mode is accessed via the control panel the Model 5100E aborts any operating mode that is act...

Page 32: ...tion gas pressure and flow rate is similar to the process gas pressure and flow rate Wait for the analyzer display to stabilize at least 5 minutes Warning Never apply an unregulated gas supply to the...

Page 33: ...lay to stabilize at least 5 minutes Warning Never apply an unregulated gas supply to the oxygen sensor Failure to do so may cause damage to the oxygen sensor 3 2 1 5 Step 5 Calibrate the analyzer Afte...

Page 34: ...then show an oxygen concentration The RUN ALM1 ALM2 and FAULT LED s will flash for 120 seconds to indicate that the analyzer is in a stabilization period This is to allow time to sweep the sample line...

Page 35: ...en it has detected that the measured oxygen concentration has exceed the set threshold value of Alarm 1 section 3 2 2 The ALM1 indicator LED will light The RUN indicator LED will remain lit The Alarm...

Page 36: ......

Page 37: ...Model 5100E control panel by pressing and holding the MODE key for at least 10 seconds until the 7 segment alphanumeric display shows to indicate that the analyzer has accessed setup mode Release the...

Page 38: ...up 4 Serial Output Format This parameter allows the user to set the RS 232 communications timed output format Valid Settings 0 Output on Request 1 Human Readable 2 Machine Code 3 Machine Code with Che...

Page 39: ...nd transmitting ASCII data packets over a standard RS 232 serial interface Access to the Serial Service Port may be made through a terminal emulator program such as HyperTerminal available in Microsof...

Page 40: ...through the service port that can be used for interfacing with the Model 5100E Standard Access ASCII dump to a PC printer or DAQ and provides basic operator access Advanced Level 1 Access Allows user...

Page 41: ...r can send a calibration value in decimal format where 100 Oxygen 1 The number format entered must be 6 decimal places 4 1 2 4 1 Disable RS 232 continuous OUTPUT ON REQUEST SSERFMT 0 While this settin...

Page 42: ...delimited data is presented in static columns the same number of columns is always transmitted in a complete data message Column headings from left to right Time since last re boot tab Mode tab O2 Con...

Page 43: ...er Setup menu U00 see Figure 14 is the Home screen in Advanced Level 1 access section 4 1 2 5 and provides access to all the parameters that may need to be adjusted by the user The interactive menu is...

Page 44: ...t of box setting The user may type Y at the prompt to initiate a restore or N and the prompt to bypass a restore This setting is accessed from the Setup Main Menu by typing F on the RS 232 Terminal To...

Page 45: ...alarm state Non Failsafe action is defined as relay coils are powered contacts are in non normal position in active alarm state The Alarm 1 and Alarm 2 Relays Failsafe setting may be set to YES or NO...

Page 46: ...evel The descending setting may be set to YES or NO This setting is accessed from the Alarm and Relay Setup Menu by typing 5 on the RS 232 terminal Refer to the Software Setup Log that is supplied by...

Page 47: ...ings on the bottom of the main CPU PCB section 4 1 3 This menu is accessed from the Analog Output Setup menu by typing 1 or J on the RS 232 terminal Refer to the Software Setup Log that is supplied by...

Page 48: ...0 0 to 100 0 This menu is accessed from the Analog Output Setup menu by typing 4 on the RS 232 terminal Contact factory before changing these settings 4 1 2 11 5 Force Maximum Voltage to O2 Concentrat...

Page 49: ...ments and can be anywhere from 0 0 to 100 0 This menu is accessed from the Analog Output Setup menu by typing 7 on the RS 232 terminal Contact factory before changing these settings 4 1 2 12 Display A...

Page 50: ...S on the RS 232 terminal To navigate backwards use the Esc or Q key on the RS 232 terminal Figure 19 Serial Output Configuration Menu 4 1 2 13 1 Baud Rate This menu sets the RS 232 baud rate The baud...

Page 51: ...terminal To navigate backwards use the Esc or Q key on the RS 232 terminal 4 1 2 15 Gas Calibration Lockout If the Gas Calibration Lockout is enabled the user may not change but only view the Gas Cali...

Page 52: ...NGE 4 1 3 Change factory settings via Hardware Jumpers 4 1 3 1 Analog Voltage Output The Analog voltage output must be configured using the hardware settings see Figure 19 In addition the software set...

Page 53: ...e Model 5100E upside down to access the jumpers through the port provided Identify the appropriate jumper position Use an insulated jumper puller to remove and replace jumpers Figure 20 4 1 3 1 3 Retu...

Page 54: ...he sensor on a known gas source should be performed during the customer s regular maintenance schedule The chart below should serve as a general guide for maintenance personnel TASK RECOMMENDED FREQUE...

Page 55: ...n Standby mode Fault 2 indicates that the analyzer relay outputs which are the alarm 1 relay alarm 2 relay fault relay and temperature OK relay are in standby mode The relays are locked at the last va...

Page 56: ...ter has not sent a new concentration measurement to the microprocessor Fault 8 is activated when the analyzer is turned on and de activated when the first Analog to Digital converter reading is made a...

Page 57: ...1 Appendix A Spare Parts List PART NUMBER DESCRIPTION 5 06 4900 08 0 Operations Manual C1 11 1220 03 0 VAC Fuses for Power Supply Board for Vac units only 1A 250 VAC Slo Blo C1 17 0052 00 0 Replacemen...

Page 58: ...C Analog Current Output Non isolated 4 20 mA 12 VDC negative ground powered by analyzer maximum electrical load 250 Ohms Range ID Voltage 0 1 0 10 0 25 0 50 0 100 5 63 VDC 0 1VDC 6 25 VDC 0 1VDC 6 88...

Page 59: ...5 65 C Sensor 23 149 F 5 65 C SAMPLE PRESSURE 15 Hg vacuum 7 PSIG 0 5 1 5 Bar SAMPLE FLOW 0 5 1 5 LPM 1 LPM nominal 1 3 cubic feet per hour POWER VAC Unit 90 264 VAC 47 63 Hz Single Phase 13 Watts VD...

Page 60: ...ixed 0 1 VDC X 0 10 Fixed 0 5 VDC 0 25 Fixed 0 10DC 0 50 Fixed 0 100 Fixed Analog Current Output Low Auto Ranging 0 20mA High Auto Ranging 4 20mA X Full Auto Ranging X Rs 232 Baud Rate Assume Low End...

Page 61: ...the control panel without entering the normal Control panel or Service Port user menus KEYS PRESSED DESCRIPTION OF FUNCTION UP DOWN Return to RUN mode from any User mode UP DOWN hold both keys for 10...

Page 62: ...0 50 0 100 5 63 VDC 6 25 VDC 6 88 VDC 7 50 VDC 8 13 VDC LOW AUTO RANGE 0 00 99 1 00 9 99 10 0 24 9 X XX X XX XX X 0 1 0 10 0 25 5 63 VDC 6 25 VDC 6 88 VDC H IGH AUTO RANGE 10 0 24 9 25 0 49 0 50 0 100...

Page 63: ...d cell Allow unit to cool down before servicing Emergency and First Aid Procedures Skin Contact with hot sensor Immediately flush skin with plenty of cold water for 15 minutes Handling Protective meas...

Page 64: ...cs reserves the right to use any materials in the manufacture repair or service of the products and to modify the design as deemed suitable in so far as these materials or modifications maintain the s...

Reviews: