Appendix B - Specifications
Manual Part Number:
C5-06-4900-08-0
Revision Level: C
ECO 8029
File Name: MN-A-0008
Page 5-3
RANGE
0
–
1 %
•
0
–
10 %
•
0
–
25 %
•
0
–
50 %
•
0
–
100 %
•
Low Auto (0
–
25 %)
•
High Auto (10
–
100 %)
•
Full Auto (0
–
100 %)
MEASUREMENT
RANGE
0-25%, 0-100%**
ACCURACY OF
MEASUREMENT
RANGE
+ .5% of measurement range
LDL
.05% @ 0-25% measurement range, .1% @ 0-100% measurement range
RESPONSE TIME
T
90
< 15 seconds
WARM UP TIME
60-seconds for sensor warm-up, 35 minutes for temperature equilibrium
of the entire analyzer.
HUMIDITY
0-95 % non-condensing
OPERATING
TEMPERATURE
Analyzer:
32-122° F (0
–
50° C)
Sensor:
32-122° F (0
–
50° C)
STORAGE
TEMPERATURE
Analyzer:
23-149° F (-5
–
65° C)
Sensor:
23-149° F (-5
–
65° C)
SAMPLE PRESSURE
15” Hg vacuum–
7 PSIG (0.5
–
1.5 Bar)
SAMPLE FLOW
0.5
–
1.5 LPM (1 LPM nominal) (1-3 cubic feet per hour)
POWER
VAC Unit:
90
–
264 VAC, 47
–
63 Hz, Single Phase, 13 Watts
VDC Unit:
11
–
18 VDC, 13 Watts
20
–
30 VDC, 13 Watts
Mains power is to be regulated to +/- 10%
MECHANICAL
Faceplate:
Height 3.75”x Width 7.00”
•
NEMA 4
•
IP66
Panel Cut-out:
Height 2.91” x Width 6.20”
Electronic
Compartment:
Height 2
.81“ x Width 5.98” x 3.60” Depth
•
NEMA 1
•
IP20
WEIGHT
3 Lbs (1.4 kg)
WARRANTY
12-months from date of shipment
**Measurement of oxygen >95% affects sensor life significantly, e.g. measuring 100% oxygen will potentially
reduce sensor life to less than 6 months. The solid state electrolyte depletes with high oxygen
concentrations.
The manufacturer reserves the right to amend these specifications as necessary.